TIENet™ 360 LaserFlow™ Velocity Sensor
Appendix C Velocity Error Codes
C-4
3: Ultrasonic Read Error
An error occurred during an ultra-
sonic reading. This includes com-
munication errors with the
ultrasonic sensor.
Ensure the blanking distance is correct. Refer to Sec-
tion 3.2 Measurement Setup.
Ensure the placement of the sensor is accurate. Refer
to Section 2.5.3 Sensor Positioning and Require-
ments.
Verify there is no foam or extreme turbulence.
4: Analog-to-Digital Con-
verter (ADC) Error
An error occurred during an ADC
reading.
Contact factory.
5: Focus Error
An error occurred while focusing
the laser.
Ensure the input voltage is 10V (minimum).
6: Digital Signal Proces-
sor (DSP) Communication
Error
An error occurred during commu-
nication with the DSP.
Contact factory.
7: Sensor not seen
Communication failed between
flow meter and sensor. This could
be due to the sensor being dis-
connected, losing power, etc.
Ensure the sensor is connected to the flow meter.
Ensure the input voltage is 10V (minimum).
Verify the battery is connected or charged.
Table C-1 Definitions of 360 Error Codes
Error Code
Meaning
Possible Solutions
Table C-2 Definitions of 350 Error Codes
Error Code
Meaning
Possible Solutions
1: Measurement
Error
Unable to generate valid velocity
data set.
Contact factory.
3: Velocity Filter Error An error occurred while setting the
filter clock.
Ensure there are uniform flow conditions at sensor
location.
5: Velocity Gain Error Unable to set gain after maximum
attempts.
Ensure there is no debris on the sensor.
6: Level to Low error
The absolute level is too low to
attempt velocity measurement.
Place sensor in flow with at least 1” deep.
Set storage of velocity to be secondary based on
depth of 1”.
7: Timeout/Quality
Error
Measurement timed out or did not
meet measurement quality
requirements.
Contact factory.
Summary of Contents for TIENet 360 LaserFlow Ex
Page 2: ......
Page 4: ......
Page 10: ...TIENet 360 LaserFlow Velocity Sensor Safety viii ...
Page 54: ...TIENet 360 LaserFlow Velocity Sensor Section 2 Preparation and Installation 2 32 ...
Page 90: ...TIENet 360 LaserFlow Velocity Sensor Section 4 Setup with 2160 LaserFlow Module 4 20 ...
Page 102: ...TIENet 360 LaserFlow Velocity Sensor Appendix A Replacement Parts A 2 A 2 LaserFlow Sensor ...
Page 103: ...TIENet 360 LaserFlow Velocity Sensor Appendix A Replacement Parts A 3 ...
Page 104: ...TIENet 360 LaserFlow Velocity Sensor Appendix A Replacement Parts A 4 ...
Page 105: ...TIENet 360 LaserFlow Velocity Sensor Appendix A Replacement Parts A 5 ...
Page 106: ...TIENet 360 LaserFlow Velocity Sensor Appendix A Replacement Parts A 6 ...
Page 107: ...TIENet 360 LaserFlow Velocity Sensor Appendix A Replacement Parts A 7 ...
Page 109: ...TIENet 360 LaserFlow Velocity Sensor Appendix A Replacement Parts A 9 ...
Page 111: ...TIENet 360 LaserFlow Velocity Sensor Appendix A Replacement Parts A 11 ...
Page 132: ...TIENet 360 LaserFlow Velocity Sensor D 6 ...
Page 136: ......
Page 138: ...Warranty ...
Page 140: ......