SeaBat T-Series Subsea Operator's Manual
Page 79
January 31, 2019
Version 7
A.2.5 Input Signal (Timing and Triggers) Errors
Message
Description
Action
Failed to register external trigger,
ERROR =
Failed to unregister external trigger,
ERROR =
FIRMWARE does not support
External Trigger. Pinging is OFF
External triggering has failed.
Check/update software installation.
External Trigger Firmware
cannot be initialized or not
supported.
Reboot the system.
Contact Teledyne
RESON Customer
Support if the problem
persists.
FATAL ERROR: PPS Timestamp
error Check 1 PPS input
The pulse is not being
detected. Time stamping will
be done using system time
only. 1 millisecond accuracy
not possible.
Check input signal
cabling.
A.2.6 Calibration Errors
Message
Description
Action
Calibration Failed: Too much noise
Calibration Failed: Over 10% dead
Rx channels.
Calibration Failed: Magnitude Error
Calibration Failed: Phase Error
Calibration Error: Calibration might
be invalid. Repeat the calibration.
Calibration has failed.
Check BITE page and
hardware and cable
installation to the
receiver. Check
environmental noise.
Contact Teledyne
RESON Customer
Support.
A.3
Warning Messages
NOTE
The warning messages listed in this section are accurate up to the document release date. For
a complete list of up-to-date warning messages, contact Teledyne RESON Customer Support
(
Warnings are marked with a yellow
in the Messages pane.
A.3.1 Configuration Warnings
Message
Description
Action
7K Bite File cannot be Read
Cannot Read UDP Configuration
file <x> Error
Cannot Open or Create UDP
Configuration file
UDP Configuration file not available
Software installation is
corrupt.
Restore factory defaults.
Reinstall software.
Summary of Contents for SeaBat T Series
Page 6: ......