Chapter 6
Operation Modes
33
•
T
xfr
(Transfer Time)
This interval represents the time it takes to transfer data from the photosites to
the vertical shift register.
This time is sensor dependent.
Refer to
•
T
read
(Readout Time)
This interval represents the time it takes to read the image from the sensor.
This time is sensor dependent.
Refer to
for Readout Times that have been measured.
•
T
r
(Recovery Time)
This interval represents the time from the end of sensor readout to the time the
system can accept another trigger pulse.
The recovery time is approximately 1
s.
•
Maximum Trigger Rate
The Maximum Trigger Rate is related to the minimum time between triggers
which may be calculated as:
T
s
+ T
int
+ T
xfr
+ T
read
+ T
r
Table 6-7: Mode 1: Typical Transfer Time (T
xfr
)
Sensor
Sensor Frequency
T
xfr
KAI-11002
30 MHz
125
s
Table 6-8: Mode 1: Typical Readout (T
read
) Times
Sensor
Taps
Sensor Frequency
Readout Time
KAI-11002
1
30 MHz
392 ms
2
30 MHz
212 ms
Summary of Contents for Princeton Instruments Nano-XF
Page 1: ...www princetoninstruments com Nano XF System Manual 4411 0116 Issue 5 October 4 2019...
Page 6: ...6 Nano XF System Manual Issue 5 This page is intentionally blank...
Page 16: ...16 Nano XF System Manual Issue 5 This page is intentionally blank...
Page 26: ...26 Nano XF System Manual Issue 5 This page is intentionally blank...
Page 38: ...38 Nano XF System Manual Issue 5 This page is intentionally blank...
Page 44: ...44 Nano XF System Manual Issue 5 This page is intentionally blank...
Page 50: ...50 Nano XF System Manual Issue 5 This page is intentionally blank...