Platinum/Optima CVE Refrigerated Sampler
3-34
3.5.9 Sample Second
Enter the number of seconds it will take to fill
the sample chamber and touch the
liquid-sensor electrodes. The sample seconds
must exceed the number of seconds to fill the
chamber. When the liquid-sensor is activated,
the timer stops and the program goes to post
purge.
3.5.10 Flow Pulses, Analog Input
Time Paced Only
–
If you are collecting
sample volumes dependent on flow (Section
3.5.11), select which type of flow rate input is
provided by the external flow measurement
device (see Section 2.8).
3.5.11 Composite Samples
Single Bottle Only
–
Use the number buttons
to type the number of composite samples to
take. Press the E
NTER
button to accept the
value.
The sampler does not calculate the maximum
number of samples. Use caution when selecting
the maximum number of samples per bottle so
they do not overfill.
• If you selected continuous in the B
OTTLE
L
IMIT
menu, the sampler collects samples
without regard for the total number of
samples.
• If you enter continuous, keep in mind
that the sampler might overfill the bottle.