Platinum/Optima CVE Refrigerated Sampler
3-10
ON
: If the sampler fails to detect liquid, it will
try again up 4 times to collect the sample.
3.3.6 Float Input
Depending if the F
LOAT
I
NPUT
options have
been factory-configured, enabling this input
will configure the sampling program to delay
sample collection until that circuit closes and
subsequently either:
•
Float Start
: continue sampling even if
the float circuit reopens,
•
Float Start/Stop
: toggle the collection of
samples on and off as the rising and
falling stream level closes and opens that
circuit. This feature may be used with
full bottle shut off for composite samples.
The sampler controller is D
ISABLED
when pins
1 and 2 are open of the external 7 pin connector
(orange wires). The controller is E
NABLED
when
pins 1 and 2 are closed.
3.3.7 Output Pins
An alarm output is based one of the following
events:
•
Program Done:
This event occurs when
a running sampling program has finished
Toggle to change
If Disable the option
Disable/ Enable
and press Enter.
is turned off.