
Section 2 Installation
2-5
CAUTION
Do not tip the sampler on its side or back. Tipping
the sampler more than 45° might cause oil to run
into the compressor inlet, which can permanently
damage the refrigeration system.
2.3 Connecting Power
WARNING
Before connecting the sampler to an AC power
source, be familiar with the Electrical
Requirements listed at the front of the Installation
and Operation manual.
The factory assembles the sampler for either
115 VAC/60 Hz, or 230 VAC/50 Hz. The serial
number label located inside the control cover
identifies the AC power configuration of your
sampler. You cannot convert the sampler in the
field.
• The mains power cord for 115 VAC
samplers is equipped with a NEMA 5-15P
plug for standard North American
outlets.
• The mains power cord for 230 VAC
samplers is equipped with an EU 1-16P
plug for standard CEE 7-7 European
outlets.
• For other types of outlets that supply
compatible AC power, convert the mains
Summary of Contents for Optima
Page 1: ...Platinum Optima CVE Pocket Guide ...
Page 2: ......
Page 4: ......
Page 18: ...Platinum Optima CVE Refrigerated Sampler 1 10 ...
Page 68: ...Platinum Optima CVE Refrigerated Sampler 3 22 3 4 4 Bottle Configuration Press F3 SEL ...
Page 106: ...Platinum Optima CVE Refrigerated Sampler 5 8 5 3 4 Sample Chamber Parts ...
Page 111: ......