3. Rotate the Voltage knob to change
the F setting to F-88 (Factory Set
Value).
Voltage
4. Use the Current knob to set the
F-88 setting to 1 (Return to factory
default settings).
Current
5. Press the Voltage knob to confirm.
ConF will be displayed when it is
configuring.
Voltage
6. Press the Function key again to
exit. The Function key light will
turn off.
Function
The F-89 configuration setting allows you to view
the T3PS version number, build date, keyboard
version, analog-control version, kernel build, test
command version and test command build date.
1. Press the Function key. The
Function key will light up.
Function
2. The display should show F-01 on the top and
the configuration setting for F-01 on the
bottom.
www.
.com
1.800.561.8187
Summary of Contents for LeCroy T3PS Series
Page 1: ...www com information itm com 1 800 561 8187 ...
Page 2: ...www com information itm com 1 800 561 8187 ...
Page 3: ...www com information itm com 1 800 561 8187 ...
Page 4: ...www com information itm com 1 800 561 8187 ...
Page 20: ...T3PS062001P T3PS40381P T3PS60251P www com information itm com 1 800 561 8187 ...
Page 30: ...www com information itm com 1 800 561 8187 ...
Page 78: ...www com information itm com 1 800 561 8187 ...
Page 79: ...www com information itm com 1 800 561 8187 ...
Page 80: ...www com information itm com 1 800 561 8187 ...
Page 81: ...www com information itm com 1 800 561 8187 ...
Page 99: ...www com information itm com 1 800 561 8187 ...
Page 100: ...www com information itm com 1 800 561 8187 ...
Page 103: ...www com information itm com 1 800 561 8187 ...
Page 135: ...2 3 4 www com information itm com 1 800 561 8187 ...
Page 136: ...6 5 7 8 9 10 T3PS_DC1 www com information itm com 1 800 561 8187 ...
Page 137: ...11 12 www com information itm com 1 800 561 8187 ...
Page 138: ...18 17 19 20 Teledyne T3PS40381P T0 02 20131205 n www com information itm com 1 800 561 8187 ...
Page 141: ... www com information itm com 1 800 561 8187 ...
Page 142: ...www com information itm com 1 800 561 8187 ...