Analog Output (Zero Pressure) (Read Only)
The Analog Output voltage or current value which is output when the pressure is at 0% of
full scale. This is controlled by S64 Product Config.
S37
Analog Output (FS Pressure) (Read Only)
The Analog Output voltage or current value which is output when the pressure is at 100% of
full scale. This is controlled by S64 Product Config.
S51
DAC Zero Code (Factory Read/Write)
The Analog Output D/A value used to represent 0% pressure. If in CALIBRATE state (SS8),
writes value to RAM and Sets the Analog Output to the level representing 0% pressure;
otherwise, writes value to RAM and FLASH
Range is 0 to 65535. Defaults are as follows:
0 to 5 V, 0 to 10 V
32764
1 to 5 V
35742
0 to 20 mA
0
4 to 20 mA
10950
S52
DAC Full Scale Code (Factory Read/Write)
The Analog Output D/A value used to represent 100% pressure. If in CALIBRATE state (SS8),
writes value to RAM and Sets the Analog Output to the level representing 100% pressure;
otherwise, writes value to RAM and FLASH
Range is 0 to 65535. Defaults are as follows:
0 to 5 V, 1 to 5V
47654
0 to 10 V
62545
0 to 20 mA, 4 to 20mA
54670
S53
Analog Output Full Scale (Factory Read/Write)
The Analog Output D/A value used to represent 100% pressure. If in CALIBRATE state (SS8),
writes value to RAM and Sets the Analog Output to the level representing 100% pressure;
otherwise, writes value to RAM and FLASH
Range is 0 to 65535. Defaults are as follows:
0 to 5 V, 1 to 5V
47654
0 to 10 V
62545
0 to 20 mA, 4 to 20mA
54670
S54
Comment (Read/Write)
This allows the end user to add up to 63 characters of free-format text for their own use.
NOTE: do not use the greater-than sign '>' in any text field because it could affect
automated command response processing since it is used as the end-of-response
character.
S62
Cal Date (Read/Factory Write)
The date the device was calibrated at the factory. This is a text field which accepts a
maximum of 63 characters.