Programming Manual
–
T3AWG3K-C Series Arbitrary Waveform Generator
True-Arb Operating Mode
56
Table 56: DIGitals:SKEW[m]
Command
DIGitals:STATe
Description
This command and query sets or returns the state of the digital
outputs.
It enables or disables the available digital channels.
Note
: if the Digital Channels number is set to ‘0’, DIGitals:STATe ON
has no effect.
Group
Output
Syntax
DIGitals:STATe {OFF|ON}
DIGitals:STATe?
Related Commands
None
Arguments
OFF|ON
•
OFF: disables the digital lines
•
ON: enables the digital lines
Returns
0|1 where ‘0’ means OFF and ‘1’ means ON
Example
DIGitals:STATe ON
Enables all the available digital channels.
SOURce1:DIGital: STATe?
Might returns “1” that means ON
Table 57: DIGitals:STATe
Syntax
DIGitals:SKEW[m] {MINimum|MAXimum|DEFault|<Seconds>}
DIGitals:SKEW[m]? [{MINimum|MAXimum}]
Related Commands
None
Arguments
•
MINimum: sets the parameter to the minimum value
•
MAXimum: sets the parameter to the maximum value
•
DEFault: sets the parameter to the default value
•
< Seconds > := <NRf> value
The value of m indicates the digital probe where m=1 Digital Pod A,
m=2 Digital Pod B, m=3 Digital Pod C, m=4 Digital Pod D
Returns
<NRf>
Example
DIGitals:SKEW1 100n
Set the skew for the digital probe A to 100n
DIGitals:SKEW1?
Might return 100.020E-9 ( The set value is automatically rounded to
the closest value that the hardware can implement).