Programming Manual
–
T3AWG3K-C Series Arbitrary Waveform Generator
True-Arb Operating Mode
46
Example
AWGCONTROL:RMODE STEPped sets the AWG run mode to
Stepped.
AWGCONTROL:RMODE? might return CONTinuous, indicating that
the AWG run mode is set to Continuous.
Table 38: AWGControl:RMODe
Command
AWGControl: RSTATe? (Query Only)
Description
This command returns the run state of the AWG.
Group
Control
Syntax
AWGControl:RSTATe?
Related Commands
AWGControl:RMODe
Arguments
Returns
A single <NR1> value.
0 indicates that the AWG has stopped.
1 indicates that the AWG is waiting for trigger.
2 indicates that the AWG is running.
Example
AWGCONTROL:RSTATe? might return 0, indicating that waveform
generation is stopped.
Table 39: AWGControl:RSTATe?
Command
AWGControl:RUN[:IMMediate] (No Query Form)
Description
This command starts the output of a sequence. This is the same
to press the run button on the front-panel or display.
Group
Control
Syntax
AWGControl:RUN[:IMMediate]
Related Commands
AWGControl:STOP[:IMMediate]
Arguments
Returns
A single <NR1> value.
0 indicates that the AWG has stopped.
1 indicates that the AWG is waiting for trigger.
2 indicates that the AWG is running.
Example
AWGControl:RUN
puts the AWG in the run state
Table 40: AWGControl:RUN[:IMMediate]