Programming Manual
–
T3AWG3K-C Series Arbitrary Waveform Generator
True-Arb Operating Mode
106
Command
SEQuence:ELEM[n]:LENGth
Description
This command sets or returns the number of samples of the waveform
for the element “n”.
Group
Sequence
Syntax
SEQuence:ELEM[n]:LENGth {MINimum|MAXimum|DEFault|
<value>}
SEQuence:ELEM[n]:LENGth? [{MINimum|MAXimum}]
Related Commands
None
Arguments
The value of n indicates the sequence element number.
<value>::=<NR1> is the length of entry n.
MINimum sets or queries the minimum value of length.
MAXimum sets or queries the maximum value of length.
DEFault sets the default value of length (2048).
Returns
<value>::=<NR1>
Example
SEQuence:ELEM3:LENGth 100
Sets the sequencer length of the element n to 100 for all channels
SEQuence:ELEM2:LENGth?
Might returns 1000.
Table 144: SEQuence:LENgth
Command
SEQuence:ELEM[n]:LOOP:COUNt
Description
This command sets or returns the number of repetitions of the
waveform
for the element “n”.
Group
Sequence
Syntax
SEQuence:ELEM[n]:LOOP:COUNt {MINimum|MAXimum|DEFault|
INFinite|<value>}
SEQuence:ELEM[n]:LOOP:COUNt? [{MINimum|MAXimum}]
Related Commands
None
Arguments
The value of n indicates the sequence element number.
<value>::=<NR1> is the number of repetitions of entry n.
MINimum sets or queries the minimum value of repetitons parameter.
MAXimum sets or queries the maximum value of repetitons
parameter.
INFinite sets infinite repetitons.
DEFault sets the default value of repetition (1).
Returns
<value>::=<NR1>
Example
SEQUENCE:ELEM1:LOOP:COUNT 120