Section 3 Preparation:
Rf 150
Systems
This section provides instructions for unpacking and installing
the Combi
Flash
Rf 150 system. To prepare the system for
operation, sequentially follow all instructions in sections 3.1
through 3.11.
Note
Section 3.13 contains an Installation Qualification checklist. If
required, sign off the checklist entries as you successfully
complete the following sections.
Note
The Rf150 is rated for operation with reagent grade solvents.
Operation at pressures > 5 bar (~75 psi) with high purity hexanes
(>99% pure) at 100% concentration may cause the pump to stall
and produce a grating sound. Many times, normal operation can
be restored by removing the high pressure load and flushing the
pump with isopropyl alcohol.
3.1 Unpacking the Unit
The Combi
Flash
Rf 150 system is shipped in a single carton.
Carefully unpack the shipment and inspect the contents.
WARNING
The system is heavy. Use a two-person lift to prevent injury.
CAUTION
Do not lift the system by the fraction collector arm. Use the lifting
handles located on the instrument side panels.
Summary of Contents for CombiFlash EZ Prep
Page 2: ...CombiFlash Rf User Manual ii ...
Page 8: ...CombiFlash Rf User Manual viii ...
Page 52: ...CombiFlash Rf User Manual 2 16 Figure 2 7 Loading test tubes ...
Page 85: ...CombiFlash Rf User Manual 3 20 ...
Page 103: ...CombiFlash Rf User Manual 4 18 ...
Page 185: ...CombiFlash Rf User Manual 6 64 ...
Page 199: ...CombiFlash Rf User Manual 7 14 ...
Page 223: ...CombiFlash Rf User Manual 8 24 ...
Page 225: ......
Page 227: ......
Page 229: ......
Page 231: ......
Page 233: ......
Page 235: ......
Page 237: ......