6712SR Refrigerated Sampler
Section 4 Standard Programming
4-9
4.6.2 Sampler
Enable/Disable
Teledyne Isco flow meters and flow loggers have a programmable
sampler-enable feature that lets them enable (start) or disable
(stop) a program according to certain monitored conditions. For
example these conditions can be level, flow rate, pH, temper-
ature, percent, rainfall, I/O; or a combination of two conditions.
The 1640 Liquid Level Actuator can also be used, as well as any
other instrument that can ground the input to pin F. Grounding
the input disables the sampler.
The sampler disregards disable signals during the delay to start
time. However, when the sampler detects a disable signal at the
start time, it suspends the program until it is enabled. The
diagram in Figure 4-2 shows the sampler’s responses.
Delayed Start Screen
1
NO DELAY TO START
DELAYED START
CLOCK TIME
WAIT FOR PHONE CALL
• Select DELAYED START.
2
FIRST SAMPLE
AFTER A
___MINUTE DELAY
(1-999)
•
Enter the number of
minutes to delay the
countdown to the first
sample.
3
START FLOW COUNT
AFTER A
__MINUTE DELAY
(1-999)
•
The sampler uses this
screen for flow paced
programs.
Clock Start Time Menu
1
NO DELAY TO START
DELAYED START
CLOCK TIME
WAIT FOR PHONE CALL
• Select CLOCK TIME.
2
FIRST SAMPLE AT
06:00
• Enter the time of day.
3
SELECT DAYS:
SU
MO TU WE TH FR
SA
DONE
•
Press an arrow key
until the cursor is on
the correct day.
Press
↵
.
•
Repeat until each day
you want is blinking.
•
Press an arrow until
DONE blinks. Press
↵
.
Summary of Contents for 6712SR
Page 2: ......
Page 4: ......
Page 14: ...6712SR Refrigerated Sampler Table of Contents xii ...
Page 28: ...6712SR Refrigerated Sampler Section 1 Introduction 1 14 ...
Page 42: ...6712SR Refrigerated Sampler Section 2 Installation Preparation 2 14 ...
Page 54: ...6712SR Refrigerated Sampler Section 3 Getting Started 3 12 ...
Page 140: ...6712SR Refrigerated Sampler Section 6 SDI 12 Sondes 6 8 ...
Page 188: ...6712SR Refrigerated Sampler Appendix A Menu Flowcharts A 24 ...
Page 196: ...6712SR Refrigerated Sampler Appendix B Material Safety Data Sheets B 8 ...
Page 206: ...6712SR Refrigerated Sampler Appendix D Replacement Parts D 2 ...
Page 207: ...6712SR Refrigerated Sampler Appendix D Replacement Parts D 3 ...
Page 208: ...6712SR Refrigerated Sampler Appendix D Replacement Parts D 4 Locknut 10 32 Screw 10 32 x 3 4 ...
Page 209: ...6712SR Refrigerated Sampler Appendix D Replacement Parts D 5 ...
Page 210: ...6712SR Refrigerated Sampler Appendix D Replacement Parts D 6 ...
Page 211: ...6712SR Refrigerated Sampler Appendix D Replacement Parts D 7 ...
Page 212: ...6712SR Refrigerated Sampler Appendix D Replacement Parts D 8 ...
Page 218: ...6712SR Refrigerated Sampler Appendix E Accessories List E 6 ...
Page 222: ...6712SR Refrigerated Sampler Index Index 4 ...
Page 224: ......