6712SR Refrigerated Sampler
Section 5 Extended Programming
5-28
Adjust Suction Head –
If you have chosen to manually enter
the suction head, and are in the manual paused state, the
scrolling menu will have an option to adjust the suction head. If
you make changes to the suction head value, the sampler logs a
changed suction head event.
5.16 Sampling Reports
The 6712 records a variety of data while running a sampling
program. Sample reports can be found in Section 4.
Section 4, Table 4-3 contains a list of source codes and program
events that may be recorded when using either standard or
extended programming.
5.17 Other Functions
In extended programming, the “Other Functions” menu screen
includes options for:
• Maintenance (see Section 7)
• Manual Functions (see Section 4)
• Software Options.
• Hardware
5.18 Software Options
The software option screens allow you to set up a number of dif-
ferent software options:
5.18.1 Liquid Detector
Enable/Disable
Under some conditions (pressurized lines, sampling foam), a
more repeatable sample volume may be delivered by disabling
the liquid detector. Selecting NO will turn the Liquid Detector
off.
If the Liquid Detector is off:
• you must enter the SUCTION HEAD
• no RINSES or RETRIES will be available
• you will not be able to CALIBRATE SAMPLE volumes
• the RESULTS report will read a COUNT TO LIQUID of
0
• A complete purge will occur between samples when
using multiple bottles-per-sample distribution.
Other Functions Menu
RUN “
PROGRAM NAME
”
PROGRAM
VIEW REPORT
OTHER FUNCTIONS
• Select OTHER FUNCTIONS.
MAINTENANCE
MANUAL FUNCTIONS
SOFTWARE OPTIONS
HARDWARE
•
Liquid Detector Enable
•
Display Backlighting
•
Normal/Quick View Screens
•
Event Marks
•
Measurement Interval
•
Pump Purge Counts
•
Dual Sampler Mode
•
Serial Output
•
Full Bottle Detection
• Remote Access
•
Interrogator Connect Power
USE LIQUID DETECTOR?
YES
NO
Summary of Contents for 6712SR
Page 2: ......
Page 4: ......
Page 14: ...6712SR Refrigerated Sampler Table of Contents xii ...
Page 28: ...6712SR Refrigerated Sampler Section 1 Introduction 1 14 ...
Page 42: ...6712SR Refrigerated Sampler Section 2 Installation Preparation 2 14 ...
Page 54: ...6712SR Refrigerated Sampler Section 3 Getting Started 3 12 ...
Page 140: ...6712SR Refrigerated Sampler Section 6 SDI 12 Sondes 6 8 ...
Page 188: ...6712SR Refrigerated Sampler Appendix A Menu Flowcharts A 24 ...
Page 196: ...6712SR Refrigerated Sampler Appendix B Material Safety Data Sheets B 8 ...
Page 206: ...6712SR Refrigerated Sampler Appendix D Replacement Parts D 2 ...
Page 207: ...6712SR Refrigerated Sampler Appendix D Replacement Parts D 3 ...
Page 208: ...6712SR Refrigerated Sampler Appendix D Replacement Parts D 4 Locknut 10 32 Screw 10 32 x 3 4 ...
Page 209: ...6712SR Refrigerated Sampler Appendix D Replacement Parts D 5 ...
Page 210: ...6712SR Refrigerated Sampler Appendix D Replacement Parts D 6 ...
Page 211: ...6712SR Refrigerated Sampler Appendix D Replacement Parts D 7 ...
Page 212: ...6712SR Refrigerated Sampler Appendix D Replacement Parts D 8 ...
Page 218: ...6712SR Refrigerated Sampler Appendix E Accessories List E 6 ...
Page 222: ...6712SR Refrigerated Sampler Index Index 4 ...
Page 224: ......