6712SR Refrigerated Sampler
Section 5 Extended Programming
5-36
When asked if you have a rain gauge connected, select YES to
record the rain gauge readings and add RAIN as a sampler
enable option. Select the appropriate tip for the rain gauge. The
sampler will then create a rain memory partition and update the
available sampler enable conditions.
Select NO to disable the readings and the sampler enable option.
When you select NO, the sampler removes the rain memory par-
tition, leaving the memory space available for other reading
types.
5.19.3 I/O Pin Programming
Pins C, H, and I on the Rain Gauge Connector are user-con-
figured as I/O1, I/O2, and I/O3 (Input/Output pins) respectively.
Each pin can be either an input for sampler enabling, or an
output for alarms.
Note
The state of the I/O lines is not defined when the sampler is
OFF.
The enable input requires a pulse signal of 5 to 15 volts. Teledyne
Isco offers optional interface modules for converting from a 4-20
mA analog or pulse duration signal.
The alarm output can be set for any RUNNING PROGRAM or
RUN ERROR state. In addition, the CONDITIONS output can
report alarm conditions using the parameters the 6712 is con-
figured to measure. You can set thresholds for one or two condi-
tions and the CONDITIONS output can then trigger alarms
when the measured parameters exceed the threshold.
The alarm output requires an alarm interface module. Teledyne
Isco offers a module designed to provide one, two, or three
contact closures. The alarm module can interface the sampler
with a large number of non-Teledyne Isco devices. For more infor-
mation about the alarm module, contact your Teledyne Isco sales
representative.
5.19.4 Refrigerator
Temperature
Teledyne Isco offers a temperature sensor intended for use with
the 6712FR Refrigerated Sampler. The optional sensor connects
to the Rain Gauge port and uses I/O3 to transmit its temperature
readings to the 6712.
Note
If alarms are being used, the temperature sensor is simply
plugged into the panel mount on the one- or two-alarm output
module. Because the temperature sensor uses I/O3, it cannot
be used with a three-alarm output module.
DO YOU HAVE A
RAIN GAuGE CONNECTED?
YES
NO
• Select YES.
RAIN GAUGE:
0.01 inch TIP
0.1 mm TIP
• The rain gauges are set up to
tip on either .01 inch or .1 mm.
Select the desired tip.
Summary of Contents for 6712SR
Page 2: ......
Page 4: ......
Page 14: ...6712SR Refrigerated Sampler Table of Contents xii ...
Page 28: ...6712SR Refrigerated Sampler Section 1 Introduction 1 14 ...
Page 42: ...6712SR Refrigerated Sampler Section 2 Installation Preparation 2 14 ...
Page 54: ...6712SR Refrigerated Sampler Section 3 Getting Started 3 12 ...
Page 140: ...6712SR Refrigerated Sampler Section 6 SDI 12 Sondes 6 8 ...
Page 188: ...6712SR Refrigerated Sampler Appendix A Menu Flowcharts A 24 ...
Page 196: ...6712SR Refrigerated Sampler Appendix B Material Safety Data Sheets B 8 ...
Page 206: ...6712SR Refrigerated Sampler Appendix D Replacement Parts D 2 ...
Page 207: ...6712SR Refrigerated Sampler Appendix D Replacement Parts D 3 ...
Page 208: ...6712SR Refrigerated Sampler Appendix D Replacement Parts D 4 Locknut 10 32 Screw 10 32 x 3 4 ...
Page 209: ...6712SR Refrigerated Sampler Appendix D Replacement Parts D 5 ...
Page 210: ...6712SR Refrigerated Sampler Appendix D Replacement Parts D 6 ...
Page 211: ...6712SR Refrigerated Sampler Appendix D Replacement Parts D 7 ...
Page 212: ...6712SR Refrigerated Sampler Appendix D Replacement Parts D 8 ...
Page 218: ...6712SR Refrigerated Sampler Appendix E Accessories List E 6 ...
Page 222: ...6712SR Refrigerated Sampler Index Index 4 ...
Page 224: ......