6712FR Refrigerated Sampler
Section 1 Introduction
1-8
Weight of Power Pack:
6.0 lbs (2.7 kg)
Weight of Battery Backed Power
Pack:
6.6 lbs (3.0 kg)
Controller Dimensions:
Length: 10.3 in (26.0 cm)
Width: 12.5 in (31.7 cm)
Height: 10 in (25.4 cm)
Temperature Ranges:
Operational: 32 to 120 °F (0 to 49 °C)
Storage: 0 to 140 °F (–18 to 60 °C)
Maximum Altitude:
2,000 Meters
Installation Category:
II
Pollution Degree:
2
Humidity:
95% RH Maximum
Enclosure:
NEMA 4X and 6
Pump: IP17
Enclosure: IP67
Typical Delivered
Volume Accuracy:
(The ability to deliver the programmed sample volume.)
±10 ml or ±10% of programmed value, whichever is greater.
Typical Repeatability:
(The ability to repeat the delivered volume for a set of samples collected under
the same conditions.)
±5 ml or ±5% of the average of the maximum and minimum sample volume in a
sample set, whichever is greater, at lifts up to 25 feet.
Typical Pump Flow Rate and
Line Transport Velocity:
Suction Head
Flow Rate
Line Transport Velocity
3 feet
3.9 liters per minute
3.0 ft/s (0.91 m/s)
5 feet
3.9 liters per minute
3.0 ft/s (0.90 m/s)
10 feet
3.7 liters per minute
2.9 ft/s (0.87 m/s)
15 feet
3.6 liters per minute
2.7 ft/s (0.83 m/s)
20 feet
3.3 liters per minute
2.5 ft/s (0.77 m/s)
23 feet
3.1 liters per minute
2.3 ft/s (0.71 m/s)
25 feet
2.8 liters per minute
2.2 ft/s (0.66 m/s)
Maximum Suction Head:
(The maximum suction head is the vertical height at which a sample can be
taken.)
28 ft (8.5 m) at 30 inHg (762 mmHg)
Pump Tube Life:
Tube life may vary depending on the abrasiveness of the sample liquid.
Recommended maximum 1,000,000 pump counts. This equates to 912 stan-
dard samples. A standard sample is 200 ml at 5 ft (1.5 m) suction head using a
10 ft (3 m) vinyl suction line.
6712 Controller: Electrical Specifications
Controller Internal Battery:
5 years minimum (maintains internal logic, program settings & stored data)
Real Time Clock Accuracy:
1 minute per month
Program Memory:
Nonvolatile programmable Flash.
Can be field updated via interrogator connector.
Sampler Power Requirements:
Nominal: 12 volts DC supplied by the refrigerator
Operational: 11 to 13 Volts DC
Absolute Maximum: 14.4 Volts DC
Power Pack Requirements:
Use only Isco-made power
packs certified by UL.
(P/N 60-1684-088)
120VAC ±12VAC, 60 Hz., 1.0 Amp.
Note: This is the
only
version certified by UL. The line cord (mains
connect cable) is the “Disconnect Device.”
Table 1-3 Technical Specifications for the 6712FR Refrigerated Sampler (Continued)
Summary of Contents for 6712FR and
Page 2: ......
Page 4: ......
Page 8: ...6712FR Refrigerated Sampler Safety vi ...
Page 16: ...6712FR Refrigerated Sampler Table of Contents viii ...
Page 56: ...6712FR Refrigerated Sampler Section 3 Programming Introduction 3 12 ...
Page 142: ...6712FR Refrigerated Sampler Section 6 SDI 12 Sondes 6 8 ...
Page 174: ...6712FR Refrigerated Sampler Section 8 Maintenance 8 22 ...
Page 198: ...6712FR Refrigerated Sampler Appendix A Menu Flowcharts A 24 ...
Page 206: ...6712FR Refrigerated Sampler Appendix B Material Safety Data Sheets B 8 ...
Page 216: ...6712FR Refrigerated Sampler Appendix C General Safety Procedures C 10 ...
Page 218: ...6712FR Refrigerated Sampler Appendix D Replacement Parts List D 2 ...
Page 219: ...6712FR Refrigerated Sampler Appendix D Replacement Parts List D 3 ...
Page 220: ...6712FR Refrigerated Sampler Appendix D Replacement Parts List D 4 ...
Page 221: ...6712FR Refrigerated Sampler Appendix D Replacement Parts List D 5 ...
Page 223: ...6712FR Refrigerated Sampler Appendix D Replacement Parts List D 7 ...
Page 224: ...6712FR Refrigerated Sampler Appendix D Replacement Parts List D 8 ...
Page 225: ...6712FR Refrigerated Sampler Appendix D Replacement Parts List D 9 ...
Page 226: ...6712FR Refrigerated Sampler Appendix D Replacement Parts List D 10 ...
Page 232: ...6712FR Refrigerated Sampler Appendix E Accessories List E 6 ...
Page 236: ...6712FR Refrigerated Sampler Index Index 4 ...
Page 238: ......
Page 240: ......
Page 242: ...Warranty ...
Page 244: ......