Section 5 Extended Programming
5-44
5.21 Pressurized Lines
The sampler can obtain samples from pressurized lines, as long
as the line pressure remains below 15 PSI (pounds per square
inch). Pressures greater than 15 PSI may prevent the sampler
from purging the suction line; moreover, extreme pressures can
force liquid past the pump, even when the pump is not running.
For more consistent sample volumes from lines under constant
pressure, calibrate the samples. Teledyne Isco does not rec-
ommend sampling from pressurized lines.
5.22 Command Driven
Operation
The 6712 controller can be used as a slave to an external con-
troller. The external controller is responsible for determining
when to take a sample, how much volume to pump, and where to
put the sample. The external controller directly interfaces to the
sampler via an RS-232 communications port at 2,400-19,200
baud, 8 data bits, no parity, and 1 stop bit. A comma-sepa-
rated-value protocol is used by the external controller to make
requests, and by the sampler to report results. Use a standard
computer connect cable (see Appendix E) to connect the sampler’s
interrogator port and the external controller.
For information on serial data codes and their meanings, refer to
Table 5-5 Serial Data Codes in this section.
The 6712 Sampler will respond to three commands: Turn
Sampler On, GET STATUS, and TAKE SAMPLE. All commands
must include the instruction set. The checksum value is optional.
The available commands are:
1. Turn Sampler On: STS,2,CS,582<CR>
• “582” is the byte-by-byte sum of the ASCII values in the
string “STS, 1,CS”.
• “<CR>” is a carriage return signifying the end of the
command.
• If the sampler is OFF, this command will turn it ON
before returning status info.
2. Get Status: STS,1,CS,581<CR>
“STS,1” instructs the sampler to return the current status.
• “CS,581” is the optional checksum value. The
byte-by-byte sum of the ASCII values in the string
“STS,1,CS” is 581.
• “<CR>” is a carriage return signifying the end of the
command.
3. Take Sample: BTL,2,SVO,100,CS,1039<CR>
The TAKE SAMPLE command is only valid if the sampler is
WAITING TO SAMPLE.
• “BTL,2” instructs the sampler to place the sample in
bottle 2.
• “SVO,100” instructs the sampler to take a 100 ml
sample.
Summary of Contents for 6712FR and
Page 2: ......
Page 4: ......
Page 8: ...6712FR Refrigerated Sampler Safety vi ...
Page 16: ...6712FR Refrigerated Sampler Table of Contents viii ...
Page 56: ...6712FR Refrigerated Sampler Section 3 Programming Introduction 3 12 ...
Page 142: ...6712FR Refrigerated Sampler Section 6 SDI 12 Sondes 6 8 ...
Page 174: ...6712FR Refrigerated Sampler Section 8 Maintenance 8 22 ...
Page 198: ...6712FR Refrigerated Sampler Appendix A Menu Flowcharts A 24 ...
Page 206: ...6712FR Refrigerated Sampler Appendix B Material Safety Data Sheets B 8 ...
Page 216: ...6712FR Refrigerated Sampler Appendix C General Safety Procedures C 10 ...
Page 218: ...6712FR Refrigerated Sampler Appendix D Replacement Parts List D 2 ...
Page 219: ...6712FR Refrigerated Sampler Appendix D Replacement Parts List D 3 ...
Page 220: ...6712FR Refrigerated Sampler Appendix D Replacement Parts List D 4 ...
Page 221: ...6712FR Refrigerated Sampler Appendix D Replacement Parts List D 5 ...
Page 223: ...6712FR Refrigerated Sampler Appendix D Replacement Parts List D 7 ...
Page 224: ...6712FR Refrigerated Sampler Appendix D Replacement Parts List D 8 ...
Page 225: ...6712FR Refrigerated Sampler Appendix D Replacement Parts List D 9 ...
Page 226: ...6712FR Refrigerated Sampler Appendix D Replacement Parts List D 10 ...
Page 232: ...6712FR Refrigerated Sampler Appendix E Accessories List E 6 ...
Page 236: ...6712FR Refrigerated Sampler Index Index 4 ...
Page 238: ......
Page 240: ......
Page 242: ...Warranty ...
Page 244: ......