5-13
conditions becomes true. The default distribution setting for
event-paced sampling is sequential (each time an event occurs,
the sampler takes one sample, placing it in one bottle). However,
any distribution setting may be used.
If you have programmed multiple conditions, and more than one
condition comes true at the same time (within your measurement
interval), one sample will be taken and all conditions that caused
the event will be noted on reports.
Table 5-3 uses
rate of change
for the event condition (see step
7). This feature enables the event condition under specific cir-
cumstances.
In the example below, a rate of change is set so that if the level
rises more than one foot in a two hour time period, the event con-
dition would be enabled and a sample taken.
In determining if the rise occurs in this example, the sampler
looks back at the previous two hour time period to see if there is
a rise of more than one foot between any valley (low point) or
peak (high point) within that time period.
Once the condition becomes true (a rise of more than one foot is
noted), the event condition is enabled. It will not be enabled
again unless the condition becomes false, and then true again,
within the two hour time window.
In event pacing, up to three different rate of change event condi-
tions can be programmed (for two-part programs, a total of six
rate of change event conditions can be defined, three for each
part).
5.9.2 Event Paced Sampling
This example uses a sampler with an attached 730 Bubbler Flow
Module, Rain Gauge, and an SDI-12 sonde at address 5 with four
parameters.
Table 5-3 Event Paced Sampling
1
UNIFORM TIME PACED
FLOW PACED
EVENT PACED
NONUNIFORM TIME
Select EVENT PACED.
2
EVENT01:
RAIN
LEVEL
FLOW
TEMP5 pH5 SP_CO
5
DO5 DONE
Select the parameters that will trigger the event. Use the arrow keys to
move through the parameters.
To define the first event for this example, select LEVEL.
3
“LEVEL” CONDITION:
SET POINT
RANGE
RATE OF CHANGE
Choose SET POINT.
4
“LEVEL” SET POINT:
1.000
ft
( 0.001 - 30.00)
Enter a set point of 1 foot.
Summary of Contents for 6712FR and
Page 2: ......
Page 4: ......
Page 8: ...6712FR Refrigerated Sampler Safety vi ...
Page 16: ...6712FR Refrigerated Sampler Table of Contents viii ...
Page 56: ...6712FR Refrigerated Sampler Section 3 Programming Introduction 3 12 ...
Page 142: ...6712FR Refrigerated Sampler Section 6 SDI 12 Sondes 6 8 ...
Page 174: ...6712FR Refrigerated Sampler Section 8 Maintenance 8 22 ...
Page 198: ...6712FR Refrigerated Sampler Appendix A Menu Flowcharts A 24 ...
Page 206: ...6712FR Refrigerated Sampler Appendix B Material Safety Data Sheets B 8 ...
Page 216: ...6712FR Refrigerated Sampler Appendix C General Safety Procedures C 10 ...
Page 218: ...6712FR Refrigerated Sampler Appendix D Replacement Parts List D 2 ...
Page 219: ...6712FR Refrigerated Sampler Appendix D Replacement Parts List D 3 ...
Page 220: ...6712FR Refrigerated Sampler Appendix D Replacement Parts List D 4 ...
Page 221: ...6712FR Refrigerated Sampler Appendix D Replacement Parts List D 5 ...
Page 223: ...6712FR Refrigerated Sampler Appendix D Replacement Parts List D 7 ...
Page 224: ...6712FR Refrigerated Sampler Appendix D Replacement Parts List D 8 ...
Page 225: ...6712FR Refrigerated Sampler Appendix D Replacement Parts List D 9 ...
Page 226: ...6712FR Refrigerated Sampler Appendix D Replacement Parts List D 10 ...
Page 232: ...6712FR Refrigerated Sampler Appendix E Accessories List E 6 ...
Page 236: ...6712FR Refrigerated Sampler Index Index 4 ...
Page 238: ......
Page 240: ......
Page 242: ...Warranty ...
Page 244: ......