6712FR Refrigerated Sampler
Section 5 Extended Programming
5-45
• “CS,1039” is the optional checksum value. The
byte-by-byte sum of the ASCII values in the string
“BTL,2,SVO,100,CS,” is 1039.
• “<CR>” is a carriage return signifying the end of the
command.
5.23 Command Driven
Sampler Responses
The 6712 Sampler’s response string is the same for each of the
three commands. A sample response string is shown below:
MO,6712,ID,2424741493,TI,35523.50000,STS,1,STI,35523.4187
5,BTL,2,SVO,100,SOR,0,CS,xxxx<CR>
Note that the sampler’s response is a series of alphabetical and
numerical pairs. The alphabetical characters are headings that
describe the following numerical value or code. A complete
description of each pair is listed below.
MO,6712, is the model number of the sampler.
ID,2424741493, is the sampler’s unique identification number.
TI,35523.50000, is the sampler’s current time.
STS,1, is the sampler’s current status. The possible states are:
1 = WAITING TO SAMPLE.
4 = POWER FAILED (for short time after power is restored).
5 = PUMP JAMMED (must be resolved before continuing).
6 = DISTRIBUTOR JAMMED (must be resolved before con-
tinuing).
9 = SAMPLER OFF.
12 = SAMPLE IN PROGRESS.
20 = INVALID COMMAND.
• identifier code is not supported.
• bottle requested is not in current configuration
• sample volume requested is outside its range (10 - 9990
ml)
• day (Set_Time) must be 5 digits and more recent than
1977
21 = CHECKSUM MISMATCH.
22 = INVALID BOTTLE. (bottle requested is not in the
current configuration)
23 = VOLUME OUT OF RANGE. ( the sample volume
requested is outside its range (10-9990 ml)
STI,35523.41875, is the most recent sample time.
BTL,2, is the bottle receiving the most recent sample.
SVO,100, is the most recent sample’s volume.
SOR,0, is the results of attempting the most recent sample. The
possible results are:
0 = SAMPLE OK.
1 = NO LIQUID FOUND.
Summary of Contents for 6712FR and
Page 2: ......
Page 4: ......
Page 8: ...6712FR Refrigerated Sampler Safety vi ...
Page 16: ...6712FR Refrigerated Sampler Table of Contents viii ...
Page 56: ...6712FR Refrigerated Sampler Section 3 Programming Introduction 3 12 ...
Page 142: ...6712FR Refrigerated Sampler Section 6 SDI 12 Sondes 6 8 ...
Page 174: ...6712FR Refrigerated Sampler Section 8 Maintenance 8 22 ...
Page 198: ...6712FR Refrigerated Sampler Appendix A Menu Flowcharts A 24 ...
Page 206: ...6712FR Refrigerated Sampler Appendix B Material Safety Data Sheets B 8 ...
Page 216: ...6712FR Refrigerated Sampler Appendix C General Safety Procedures C 10 ...
Page 218: ...6712FR Refrigerated Sampler Appendix D Replacement Parts List D 2 ...
Page 219: ...6712FR Refrigerated Sampler Appendix D Replacement Parts List D 3 ...
Page 220: ...6712FR Refrigerated Sampler Appendix D Replacement Parts List D 4 ...
Page 221: ...6712FR Refrigerated Sampler Appendix D Replacement Parts List D 5 ...
Page 223: ...6712FR Refrigerated Sampler Appendix D Replacement Parts List D 7 ...
Page 224: ...6712FR Refrigerated Sampler Appendix D Replacement Parts List D 8 ...
Page 225: ...6712FR Refrigerated Sampler Appendix D Replacement Parts List D 9 ...
Page 226: ...6712FR Refrigerated Sampler Appendix D Replacement Parts List D 10 ...
Page 232: ...6712FR Refrigerated Sampler Appendix E Accessories List E 6 ...
Page 236: ...6712FR Refrigerated Sampler Index Index 4 ...
Page 238: ......
Page 240: ......
Page 242: ...Warranty ...
Page 244: ......