64
Appendix
A-1 Specifications
Packaging:
Explosion-proof.
Bulkhead mount.
Sensor:
L-2C trace analysis Micro-Fuel Cell.
Cell Block:
Application dependent.
Ranges:
Three user definable ranges from 0-100
ppm to 0-250,000 ppm, plus air calibration
range of 0-250,000 (25%).
Autoranging with range ID output.
Sample System:
Flow indicator visible from front of unit.
Positive pressure service.
Vacuum service (optional).
Auto Cal / Auto Zero. (Available with
optional, electrically operated valves.)
Alarms:
One system-failure alarm contact to detect
power failure.
Two adjustable concentration threshold
alarms with fully programmable setpoints.
Diagnostics:
Start-up or on-demand self-testing
function.
Displays:
2 line by 20 alphanumeric, VFD screen,
and one 5 digit LED display.
Digital Interface:
Full duplex RS-232 communications port.
Power:
Universal power supply 115 or 230 V ac,
at 50 or 60 Hz.
Operating Temperature:
40°C