Maintenance Model
3020T
Teledyne
Analytical
Instruments
58
b)
Check for leaks downstream from the cell, where oxygen
may be leaking into the system.
If there are no leaks and the span gas is OK, replace the cell.
5.3.3 Removing the Micro-Fuel Cell
The Micro-Fuel cell is located inside the cell block behind the front
panel (see Figure 5-1). To remove an existing cell:
1.
Remove power to the instrument by unplugging the power
cord at the power source.
2.
Open the front panel door by pressing the release button on
the top right corner of the door all the way in with a narrow
gauge tool, such as a small screwdriver, and releasing it.
3.
Leave the cell block installed. With one hand placed
underneath the cell block ready to catch the Micro-Fuel cell,
lift up on the gate in front of the cell block. This releases the
cell and cell holder from the block. The cell and holder will
fall out in your hand.