Trace Oxygen Analyzer
Installation
Teledyne
Analytical
Instruments
19
Installation
Installation of the Model 3020T Analyzer includes:
1. Unpacking
2. Mounting
3. Gas connections
4. Electrical connections
5. Installing the Micro-fuel Cell
6. Testing the system.
3.1 Unpacking the Analyzer
The analyzer is shipped with all the materials you need to install
and prepare the system for operation. Carefully unpack the analyzer and
inspect it for damage. Immediately report any damage to the shipping
agent.
3.2 Mounting the Analyzer
The Model 3020T is designed for bulkhead mounting in hazardous
environments. There are four mounting lugs—one in each corner of the
enclosure, as shown in Figure 3-1. The outline drawing, at the back of
this manual, gives the mounting hole size and spacing. The drawing also
contains the overall dimensions. Do not forget to allow an extra 1
3
/
8
" for
the hinges.
Be sure to allow enough space in front of the enclosure to swing the
door open—a 16
1
/
4
" radius, as shown in Figure 3-2.
All electrical connections are made via cables which enter the
explosion-proof housing through ports in its side. No conduit fittings are
supplied. The installer must provide two 3/4" NPT and two 1" NPT
adapters and the appropriate sealing conduit.