3 Installation
Model 3010TB
3-6: Part I
Teledyne Analytical Instruments
Voltage Signal
Current Signal
%O2
Output (V dc)
Output (mA dc)
0
0.0
4.0
1
0.1
5.6
2
0.2
7.2
3
0.3
8.8
4
0.4
10.4
5
0.5
12.0
6
0.6
13.6
7
0.7
15.2
8
0.8
16.8
9
0.9
18.4
10
1.0
20.0
To provide an indication of the range, a second pair of analog output
terminals are used. They generate a steady preset voltage (or current when
using the current outputs) to represent a particular range. Table 3-2 gives the
range ID output for each analysis range.
Table 3-2: Analog Range ID Output - Example
Range
Voltage (V)
Current (mA)
LO
0.25
8
MED
0.50
12
HI
0.75
16
CAL (0-25%)
1.00
20
Alarm Relays:
Summary of Contents for 3010TB
Page 8: ...viii Model 3010TB Teledyne Analytical Instruments...
Page 12: ...iv Part I Model 3010TB Oxygen Analyzer Teledyne Analytical Instruments...
Page 16: ...2 4 Part I 2 Operational Theory Model 3010TB Teledyne Analytical Instruments...
Page 58: ...iv Part II Model 3010TB Oxygen Analyzer Teledyne Analytical Instruments...
Page 62: ...1 4 Part II 1 Introduction Model 3010T Teledyne Analytical Instruments...
Page 98: ...A 10 Appendix Model 3010TB Teledyne Analytical Instruments...