Part 1 Control Unit
Model 3010MA
Teledyne Analytical Instruments
60
2. To change between American and European fuses, remove the
single retaining screw, flip Fuse Block over 180 degrees, and
replace screw.
3. Replace fuse as shown in Figu re 5-2.
4. Reassemble Housing as shown in Figure 5-1.
American Fuses
European Fuses
Figure 5-2: Installing Fuses
5.3 System Self Diagnostic Test
1. Press the
System
button to enter the system mode.
2. Use the < > arrow keys to move to
More
, and press
Enter.
3. Use the < > arrow keys to move to
Self-Test
, and press
Enter.
The following failure codes apply:
Table 5-1: Self-Test Failure Codes
Power
0 OK
1 5
V
Failure
2 15
V
Failure
3 Both
failed
Analog
0 OK
1
D/A A (0-1 V Concentration)
2
D/A B (0-1 V Range ID)
3 Both
Failed