WaveSurfer 3000 Oscilloscopes
Memory
The oscilloscope is equipped with two internal memory slots (M
x
) to which you can copy any channel, math, zoom or
other special waveform that is active on the grid. This is a convenient way to store an acquisition for later viewing and
analysis.
Memories are created at the same scale as the source trace, but they can be rescaled independent of the original by
using the
that appear next to the M
x
dialogs.
Trace (.trc) files saved on other Teledyne LeCroy oscilloscopes can also be stored to the removable MicroSD card. Use
the
function to save external files to memory.
Save Waveform to Memory
1. With the source waveform displayed on the grid, press the front panel
Mem button
or choose
Math > Memory
Setup
to open the Memories dialog.
2. Touch the
M
x
tab
corresponding to the memory slot you wish to use.
NOTE
: Try to choose an empty slot, as anything currently stored in that location will be overwritten. All memories will
state if they are empty or an acquisition is stored there.
3. In
Copy from Waveform
, choose the source trace to copy to memory.
4. Optionally, check
Trace On
to immediately display the memory after copying. The memory will open in the same grid
as the source trace.
5. Touch
Copy Now
.
6. Use the Zoom controls to rescale the memory when On.
Restore Memory
1. Access the Memories dialog by pressing the front panel
Mem button
or choosing
Math > Memory Setup
.
2. Check
On
next to the memory you wish to display. A description of the memory showing the source channel and
creation time appears next to each M
x
on the dialog.
58