Operator's Manual
Rescaling and Assigning Units
This feature allows you to apply a multiplication factor (
a
) and additive constant (
b
) to your waveform:
a
X +
b
. You can do
it in the unit of your choice, depending on the type of application.
Set Up Rescaling
1. Follow the usual steps to
, selecting
Rescale
from the
Functions
submenu.
2. Touch the
Rescale
right-hand dialog tab.
3. To apply a multiplication factor:
l
Check the
First multiply by:
box and enter a value for
a
, the multiplication factor.
l
Touch
then add:
and enter a value for
b
, the additive constant.
4. To change the output unit of measure from that of the source waveform:
l
Check
Override units
.
l
In
Output
enter the abbreviation for the unit the measure you wish to use.
You can also enter combinations of the unit abbreviations following these rules:
l
For the quotient of two units, use the character "
:/
"
l
For the product of two units, use the character "
.
"
l
For exponents, append the digit to the unit without a space: S2 = seconds squared.
NOTE
: Some units may be converted to simple units (e.g., V.A will display as W).
923648 Rev A
53