46
Teledyne API Model 480L O
3
Monitor User Manual
09419A DCN8342
Holding Registers (Read/Write)
Register
Address
Description
Tag Name
Data Type
#
Bits
6
Stream 4 Monitoring Duration
STREAM_4_DURATION
Integer
32
8
Stream 5 Monitoring Duration
STREAM_5_DURATION
Integer
32
10
Stream 6 Monitoring Duration
STREAM_6_DURATION
Integer
32
12
Stream 1 HI Alarm Limit (PPB) STREAM_1_HI_ALARM_SETPOINT
Inverse
Float(Double) 32
14
Stream 2 HI Alarm Limit (PPB) STREAM_2_HI_ALARM_SETPOINT
Inverse
Float(Double) 32
16
Stream 3 HI Alarm Limit (PPB) STREAM_3_HI_ALARM_SETPOINT
Inverse
Float(Double) 32
18
Stream 4 HI Alarm Limit (PPB) STREAM_4_HI_ALARM_SETPOINT
Inverse
Float(Double) 32
20
Stream 5 HI Alarm Limit (PPB) STREAM_5_HI_ALARM_SETPOINT
Inverse
Float(Double) 32
22
Stream 6 HI Alarm Limit (PPB) STREAM_6_HI_ALARM_SETPOINT
Inverse
Float(Double) 32
24
Stream 1 HI-HI Alarm Limit
(PPB)
STREAM_1_HI_HI_ALARM_SETPOINT
Inverse
Float(Double) 32
26
Stream 2 HI-HI Alarm Limit
(PPB)
STREAM_2_HI_HI_ALARM_SETPOINT
Inverse
Float(Double) 32
28
Stream 3 HI-HI Alarm Limit
(PPB)
STREAM_3_HI_HI_ALARM_SETPOINT
Inverse
Float(Double) 32
30
Stream 4 HI-HI Alarm Limit
(PPB)
STREAM_4_HI_HI_ALARM_SETPOINT
Inverse
Float(Double) 32
32
Stream 5 HI-HI Alarm Limit
(PPB)
STREAM_5_HI_HI_ALARM_SETPOINT
Inverse
Float(Double) 32
34
Stream 6 HI-HI Alarm Limit
(PPB)
STREAM_6_HI_HI_ALARM_SETPOINT
Inverse
Float(Double) 32
36
Stream 1 Analog Range (PPB) STREAM_1_RANG
Inverse
Float(Double) 32
38
Stream 2 Analog Range (PPB) STREAM_2_RANG
Inverse
Float(Double) 32
40
Stream 3 Analog Range (PPB) STREAM_3_RANG
Inverse
Float(Double) 32
42
Stream 4 Analog Range (PPB) STREAM_4_RANG
Inverse
Float(Double) 32
44
Stream 5 Analog Range (PPB) STREAM_5_RANG
Inverse
Float(Double) 32
46
Stream 6 Analog Range (PPB) STREAM_6_RANG
Inverse
Float(Double) 32