Removal and Replacement Procedures
Industry-acceptable lead-free solder or solder paste. (See Table 6.)
Table 6: Lead-free solder and paste
Material
Alloy
Vendor
Detail
Wire Core Solder
SAC305
Kester
275 w/Pb-free alloy
Syringe Paste
SAC305
Kester R
276SR w/Pb-free
alloy
Soldering iron (and tip) that provides a higher temperature (700 °F) for proper
melting of the lead-free solder.
If you are using syringe paste, a hot air gun capable of providing temperatures
slightly higher than 700 ºF.
Preparation for Module
Removal
Equipment Required.
Most modules in the instrument can be removed with a
screwdriver handle mounted with a size T-10 or T-15 Torx® screwdriver tip. All
equipment required to remove and reinstall the modules is listed. (See Table 7
on page 23.)
22
WVR4000 and WVR5000 Service Manual