Specifications
VX1411 MainFrame Instruction Manual
4–7
Table 4–5: Certifications and compliances
Characteristic
Description
EC Declaration of Conformity
Meets intent of Directive 89/336/EEC for Electromagnetic
Compatibility and Low Voltage Directive 73/23/ECC for Product
Safety. Compliance was demonstrated to the following
specifications as listed in the Official Journal of the European
Communities:
EMC Directive 89/336/EEC:
EN 55011
Class A Radiated and Conductive Emissions
EN60555-2
AC Power Line Harmonic Emissions.
EN 50082-1 Immunity:
IEC 801-2
Electrostatic Discharge Immunity
IEC 801-3
RF Electromagnetic Field Immunity
IEC 801-4
Electrical Fast Transient/Burst Immunity
IEC 801-5
Power Line Surge Immunity
Low Voltage Directive 73/23/EEC:
EN61010-1
Safety requirements for electrical equipment
for measurement, control, and laboratory use
Immunity, Enclosure,
Electrostatic Discharge
(ESD)
Up to 8 kV with no change to control settings or impairment of
normal operation, and up to 15 kV with no loss of stored data or
damage that prevents recovery of normal operation by the user.
(IEC 801-2)
Immunity, Enclosure, Radio
Frequency Electromagnetic
Field
No mainframe failures when the mainframe is subjected to a
3 V/m electromagnetic field over the frequency range of 27 MHz
to 500 MHz. (IEC 801-3)
Immunity, Fast Transients,
Common Mode
No loss of stored data, change to control settings, degradation
of performance, or temporary loss of function will occur when the
mainframe is subjected to the following transients: (IEC 801-4)
Port
Peak Voltage
Tr/Th
Rep Frequency
Signal & Control
0.5 kV
5/50 ns
5 kHz
DC Power
0.5 kV
5/50 ns
5 kHz
AC Power
1.0 kV
5/50 ns
5 kHz
Immunity, AC Power Line
Transients
No loss of stored data, change to control settings, degradation
of performance, or temporary loss of function will occur when the
mainframe is subjected to the transients as described below.
(IEC 801-5)
Mode
Peak Voltage Tr/Th
1
Common
2 kV
1.2/50
s
Differential
1 kV
8/20
s
FCC
The mainframe complies with the radiated and conducted
emissions requirements of FCC 47 CFR, Part 15, Subpart B,
Class A equipment
1
Open circuit voltage (short circuit current) transient characteristics.
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com