POSITION RANGE ADJ. R5870

mid-range

LR5804

slug above winding

LR5812

slug above winding

HP PEAKING

mid-scale

Plug R unit into the oscilloscope through flexible plug-in extension and

 

quickly check the transistorized supplies.  (Typical readings)

Check points

Knobs at preset location

 

unless otherwise speoified

Collector of V6837

 

to ground

COLLECTOR SOURCE VOLTS

 

full left (ccw) 

full right (ow)

 

-22 v 

-8.8 v

Floating chassis to

 

collector of V6857

PULSE VOLTS VARIABLE

 

full left (ocw) 

full right (cw)

 

-4.2 v 

-10.8 v

Floating chassis to

 

collector of V6877

-10.4 v

1. 

CHECK DC OUTPUT LEVEL

Measure between, pin 1 and ground,  then between pin 3 and ground of the

 

16-pin amphenol plug (65-70 v).

2. 

CHECK MICROPHONICS

Rap lightly with the palm of the hand on top of the plug-in to check

 

microphonics.

3. 

ADJUST POSITION RANGE ADJ.  (R5807)

With the front panel POSITION control centered, adjust R5807 to

 

position the trace on the "Vertical-System Electrical Center".

4. 

SET GAIN ADJUST (R5885)

Connect 

200 

mv of CALIBRATOR signal to SCOPE INPUT.  Depress PUSH TO

 

CONNECT SCOPE INPUT TO THIS TERMINAL and adjust R5885 for 

cm of

 

vertical deflection.

5. 

CHECK COLLECTOR SOURCE VOLTS FOR TRACKING

Connect voltmeter from C_ to ground and check COLLECTOR SOURCE VOLTS

 

control for tracking with front panel indication from 1_ to 1£ volts

 

on both -PNP and +NPN.

6. 

SET PULSE VOLTS ADJ. 

(R6859)

Connect vertically-calibrated test scope to JB and observe amplitude

 

of pulses generated by the mercury pulser.  Adjust R6859 for 10 volts

 

of signal output.  Notes  BIAS VOLTS must be set for 0 volts across

 

C5802 to insure accuracy of this adjustment.

R-3

January,  195>9

Summary of Contents for TYPE R

Page 1: ...s are closer than advertised specifications This is done pur posely in order to insure that the instrument will meet or exceed advertised specifications when it reaches the customer These calibration procedures and test specifications should be used therefore as a guide only Some of the test equipment referred to in the calibration procedures is not available commercially the Tektronix field engi ...

Page 2: ... HORIZONTAL DISPLAY TRIGGERING MODE TRIGGER SLOPE STABILITY TIME CM MULTIPLIER TYPE 107 INTERNAL SWEEP Type 5 Ul MAIN SWEEP NORMAL Type k AC SLOW INT full right cw 1 pSEC 1 Use a 2 2 cable terminated with a TERMINATING RESISTOR B 2 R The Vertical System Electrical Center of the 5 0 oeries oscilloscope should be determined in the following manner Using a TEST LOAD UNIT depress the PRESS TO SHORT IN...

Page 3: ...t 7 tt 8 It 9 10 11 12 13 5 K 1 9 K 6 5 K Infinite tt Ik ft 15 16 75 a Infinite Bridge out the resistors on the SERIES RESISTOR switch and the COLLECTOR CURRENT MA CM switch to insure that they are within 2 of their rated values PRESET CONTROLS COLLECTOR SOURCE VOLTS NPN COLLECTOR CURRENT MA CM 5 200 q POSITION mid range PULSE VOLTS 10 PULSE VOLTS VARIABLE full right cw TRIGGER SINGLE BIAS VOLTS 0...

Page 4: ...y with the palm of the hand on top of the plug in to check microphonics 3 ADJUST POSITION RANGE ADJ R5807 With the front panel POSITION control centered adjust R5807 to position the trace on the Vertical System Electrical Center 4 SET GAIN ADJUST R5885 Connect 200 mv of CALIBRATOR signal to SCOPE INPUT Depress PUSH TO CONNECT SCOPE INPUT TO THIS TERMINAL and adjust R5885 for 2 cm of vertical defle...

Page 5: ...ads in series across C68U7 Turn COLLECTOR SOURCE VOLTS slowly to the right cw and check to see that power supply voltage remains the same and stays in regulation as COLLECTOR SOURCE VOLTS is varied throughout its range with load and without load 10 ADJUST HF PEAKING R6808 Install TYPE R unit in plug in compartment Turn TRIGGER SLOPE to INT Apply square wave from 107 to SCOPE INPUT through special ...

Page 6: ...ere the waveform intersects these graticule lines and convert to mpsec This checks the risetime of the entire system i e pulser R unit vertical amplifier and the 5itO Series oscilloscope Approxi mately 12 mpsec Note For accurate risetime measurements check timing on the portion of the sweep used for these measurements with 50 me sine waves from an accurate source 13 CHECK SINGLE AND DOUBLE TRIGGER...

Reviews: