background image

7. 

CHECK PULSE VOLTS SWITCH STEPS

Switch PULSE VOLTS  switch to 5 and check for 5 volts of signal 

amplitude.  Check each step of this switch for indicated output 

within ±2$.  Check VARIABLE PULSE VOLTS  control to be  smooth in 

operation and. tvpve a range from 1 to 2,5! times attenuation,

8. 

CHECK USAS VOLTS

Connect meter across C58Q2 and position 

BIAS VOLTS 

control to 

obtain

 

0

 volts reading. 

Index knob to 

0

 and tighten set screw.  Check con­

trol for tracking with front panel indication throughout its range, 

on both XI and X10,

9. 

CHECK REGULATION OF ALL POWER SUPPLIES

Set  BIAS VOLTS at 0 and XI,  Attach test probe and voltmeter to the

 

output of the  bias power supply.  Connect the 5-8 test load across

 

C5>802.  Turn BIAS VOLTS  slowly to the right  (cw) and observe the volt­

age at which the  power supply goes out of regulation  (typically 3»3 v).

Turn 

VOLTS VARIABLE full left  (ccw).  Attach test probe to the

output of the floating power supply.  Connect the voltmeter and the

 

5-8 

test load across 

C5801. 

Turn PULSE VOLTS VARIABLE slowly to the

 

right  (cw)  and observe  the v  oltage at which the power supply goes out

 

of regulation  (typically i

j

.

 

v ) ,

Turn 

COLLECTOR 

SOURCE 

VOLTS full left  (ccw).  Attach test probe and 

meter to  the 

output 

of the collector power supply.  Connect 5 and 25-8 

test loads 

in 

series across C68U7.  Turn COLLECTOR SOURCE VOLTS slowly 

to  the right  (cw)  and check to see  that power supply voltage remains the 

same  and stays 

in 

regulation as COLLECTOR SOURCE VOLTS is varied 

throughout its range with load and without load,

10. 

ADJUST HF PEAKING  (R6808)

Install TYPE R unit in plug-in compartment.  Turn TRIGGER SLOPE to 

+INT.  Apply square wave  from 107 to SCOPE INPUT through special 

scope input adapter.  Depress PUSH TO CONNECT SCOPE INPUT TO THIS 

TERMINAL.  With TIME/CM at 1 pSEC X2,  obtain a  stable display through 

adjustment of TRIGGER LEVEL and STABILITY controls.  Obtain approxi­

mately 3 cm of vertical deflection and adjust R6808 for optimum 

leading edge.

11. 

ADJUST PULSER HF COMPENSATIONS  (L$80)4 and L5812)

Remove TYPE R plug-in from plug-in box and insert short extension. 

Re-install plug-in.  Connect ^TRIGGER cable to TRIGGER IN and turn 

TRIGGER SLOPE to  +EXT.  Turn COLLECTOR SOURCE VOLTS  to PNP.  Turn 

PULSE VOLTS to - , T v o l t s  and CALIBRATED.  Turn BIAS VOLTS full 

right TcwF*and IloT  Turn COLLECTOR CURRENT MA/CM to  .5 and SERIES 
RESISTOR to 50.  Turn TIME7cFTto 7 1  pSEC/CM and STABILITY full right 

Tcw7.  Install grounded emitter transistor board using 

2

iJjpldi

 transistor 

on the  front of the  R unit.  Turn BIAS VOLTS slowly to the left  (ccw) 

until approximately 3  cm of vertical deflection are  obtained.  Adjust 

STABILITY and TRIGGER LEVEL to obtain a stable display.  Then adjust 

158011 and L5812  for optimum square-wave presentation.

January, 

\

9%9

R-U

Summary of Contents for TYPE R

Page 1: ...s are closer than advertised specifications This is done pur posely in order to insure that the instrument will meet or exceed advertised specifications when it reaches the customer These calibration procedures and test specifications should be used therefore as a guide only Some of the test equipment referred to in the calibration procedures is not available commercially the Tektronix field engi ...

Page 2: ... HORIZONTAL DISPLAY TRIGGERING MODE TRIGGER SLOPE STABILITY TIME CM MULTIPLIER TYPE 107 INTERNAL SWEEP Type 5 Ul MAIN SWEEP NORMAL Type k AC SLOW INT full right cw 1 pSEC 1 Use a 2 2 cable terminated with a TERMINATING RESISTOR B 2 R The Vertical System Electrical Center of the 5 0 oeries oscilloscope should be determined in the following manner Using a TEST LOAD UNIT depress the PRESS TO SHORT IN...

Page 3: ...t 7 tt 8 It 9 10 11 12 13 5 K 1 9 K 6 5 K Infinite tt Ik ft 15 16 75 a Infinite Bridge out the resistors on the SERIES RESISTOR switch and the COLLECTOR CURRENT MA CM switch to insure that they are within 2 of their rated values PRESET CONTROLS COLLECTOR SOURCE VOLTS NPN COLLECTOR CURRENT MA CM 5 200 q POSITION mid range PULSE VOLTS 10 PULSE VOLTS VARIABLE full right cw TRIGGER SINGLE BIAS VOLTS 0...

Page 4: ...y with the palm of the hand on top of the plug in to check microphonics 3 ADJUST POSITION RANGE ADJ R5807 With the front panel POSITION control centered adjust R5807 to position the trace on the Vertical System Electrical Center 4 SET GAIN ADJUST R5885 Connect 200 mv of CALIBRATOR signal to SCOPE INPUT Depress PUSH TO CONNECT SCOPE INPUT TO THIS TERMINAL and adjust R5885 for 2 cm of vertical defle...

Page 5: ...ads in series across C68U7 Turn COLLECTOR SOURCE VOLTS slowly to the right cw and check to see that power supply voltage remains the same and stays in regulation as COLLECTOR SOURCE VOLTS is varied throughout its range with load and without load 10 ADJUST HF PEAKING R6808 Install TYPE R unit in plug in compartment Turn TRIGGER SLOPE to INT Apply square wave from 107 to SCOPE INPUT through special ...

Page 6: ...ere the waveform intersects these graticule lines and convert to mpsec This checks the risetime of the entire system i e pulser R unit vertical amplifier and the 5itO Series oscilloscope Approxi mately 12 mpsec Note For accurate risetime measurements check timing on the portion of the sweep used for these measurements with 50 me sine waves from an accurate source 13 CHECK SINGLE AND DOUBLE TRIGGER...