Functional Check and Calibration
7.
In the menu bar, select Vertical and then
select Probe Cal.
The Probe Setup dialog box appears. (Some
oscilloscopes support an automated TriMode
calibration routine and switch automatically
between input modes as shown below.)
8.
Select Clear ProbeCal, and then select
Calibrate Probe.
The probe calibration routine runs, optimizing
the probe to the oscilloscope for both probe
attenuation settings. The displayed results
differ between oscilloscope models:
9.
Oscilloscopes that do not support the
automated TriMode calibration routine
display
Pass
in the Probe Status box.
Repeat steps 5 through 8 for the remaining
input mode settings.
NOTE.
If the Probe Cal routine fails, check the
connection at the probe body. (See page 5,
Connecting Accessories to the Probe Body.)
18
P7504 & P7506 TriMode Probes Quick Start User Manual