Kit Description
TLA7012 Portable Logic Analyzer Rackmount Kit Instructions
7
When the instrument is installed according to the instructions in this document,
the rackmounted instrument meets all warranted requirements listed in the
instrument specification except for those listed in
Environmental Requirements
on page 8. Instruments mounted using methods other than those described in
these instructions may not meet their warranted requirements.
See
Specifications
in the user or service manual that applies to your instrument
model for tables of the warranted characteristics.
Cooling air enters from the bottom-front and right sides as shown in Figure 3.
You must provide adequate cool air to meet the ambient temperature require-
ments listed in Table 1.
Hot air out (rear)
Measure
input air
temperature
Figure 3: Instrument cooling
Warranted Characteristics