![Tektronix TLA7012 Instructions Manual Download Page 10](http://html1.mh-extra.com/html/tektronix/tla7012/tla7012_instructions-manual_1078222010.webp)
Kit Description
8
TLA7012 Portable Logic Analyzer Rackmount Kit Instructions
The following environmental characteristics supercede those listed in the user or
service manual for your instrument.
Table 1: Warranted characteristics
Characteristic
Description
Temperature, Inside Rack Cabinet
Operating
+10
_
C to +45
_
C (+50
_
F to +113
_
F)
Non-operating
--22
_
C to +60
_
C (-- 7.6
_
F to +140
_
F)
Clearance Requirements
The rack in which the rack adapted instrument is mounted must provide the
following clearance requirements:
H
A minimum of 14.0 inches (355.6 mm) of vertical space.
H
A minimum width of 17.717 inches (450 mm) between the left- and
right-front rails in the rack.
H
A minimum depth of 20 inches (508 mm).
CAUTION.
Adhering to these clearance requirements provides the rack-mounted
instrument with sufficient clearance for air circulation and accommodation of
the power cord and mounting hardware. Failure to provide these clearances can
result in overheating and can cause instrument faults or failure.
Environmental
Requirements