![Tektronix TLA500 Series Technical Reference Download Page 36](http://html1.mh-extra.com/html/tektronix/tla500-series/tla500-series_technical-reference_1078207036.webp)
Performance Veri
fi
cation Procedures
Test Equipment
These procedures use external, traceable signal sources to directly test
characteristics that are designated as checked
in the
Speci
fi
cations
chapter of
this manual. Always warm up the equipment for 30 minutes before beginning
the procedures.
Table 22: Test equipment
Item number and
description
Minimum requirements
Example
1.
Logic analyzer
TLA5000 or TLA5000B series logic analyzer
TLA5201, TLA501B, TLA5202,
TLA5202B, TLA5203, TLA5203B,
TLA5204, TLA5204B
2.
Logic analyzer probes
General purpose Tektronix logic analyzer probes; one probe
required for every 17 channels.
P6410, P6417, or P6418
3.
Frequency counter
Frequency accuracy: <0.0025% Frequency range: 1 kHz to
100 MHz
Agilent 33131A
4.
Digital timing
generator
1
>235 MHz, ± 0.1% accuracy
DC output: 0–5 V, ± 0.1% accuracy
Tektronix DTG5274
5.
DC voltage source
(see footnote)
0–5 V, ± 0.1% accuracy
Tektronix PS281
6.
Digital multimeter with
probes
2
6.5 digit display, 35 ppm, 1 year accuracy, 1000 readings per
minute
Fluke 8845A/8846A
7.
Cable, precision
50
Ω
coaxial
50
Ω
, 36 in, male-to-male BNC connectors
Tektronix part number 012-0482-XX
8.
Setup and Hold test
fi
xture
User-built (See page 34,
9.
Threshold Accuracy
test
fi
xture
User-built (See page 34,
1
Some timing generators (for example, the Tektronix DTG5274) include an internal DC voltage source that can be used instead of a separate power source.
2
Only needed to verify the output of the voltage source if the source does not meet speci
fi
cation. In this case, the DMM becomes the traceable instrument.
22
TLA5000 Series Product Speci
fi
cations & Performance Veri
fi
cation
Summary of Contents for TLA500 Series