Command Groups
2–60
TLA 720 Benchtop Chassis and TLA 7XM Expansion Chassis Service Manual
STATus:QUEStionable:VOLTage?
Returns the contents of the Voltage Event register. A value of one in a bit
indicates the corresponding power supply failure event has been asserted.
Unused bits always return a value of zero. The register is cleared by the *CLS
command.
There is no command associated with this query.
STATus:QUEStionable:VOLTage[:EVENt]?
Parameter
Query response
Not applicable
<NR1>
Voltage status word bits
Definition
D0
+24 V
D1
+12 V
D2
+5 V
D3
+5 V Standby
D4
+5 V External
D5
–2 V
D6
–5.2 V
D7
–12 V
D8
–24 V
D9
Not Used
D10
Not Used
D11
Not Used
D12
Not Used
D13
Not Used
D14
Not Used
D15
Not Used
Not Applicable
None
Syntax
Parameters
Reset Value
Errors and Events
Summary of Contents for TLA 720
Page 13: ...Table of Contents viii TLA 720 Benchtop Chassis and TLA 7XM Expansion Chassis Service Manual...
Page 22: ...Specifications...
Page 23: ......
Page 38: ...Operating Information...
Page 39: ......
Page 105: ...Command Groups 2 66 TLA 720 Benchtop Chassis and TLA 7XM Expansion Chassis Service Manual...
Page 111: ...TEST Subsystem 2 72 TLA 720 Benchtop Chassis and TLA 7XM Expansion Chassis Service Manual...
Page 124: ...Theory of Operation...
Page 125: ......
Page 128: ...Maintenance...
Page 129: ......
Page 155: ...Jumper Settings 4 26 TLA 720 Benchtop Chassis and TLA 7XM Expansion Chassis Service Manual...
Page 178: ...Options...
Page 179: ......
Page 182: ...Diagrams...
Page 183: ......
Page 186: ...Replaceable Mechanical Parts...
Page 187: ......
Page 198: ...Replaceable Parts TLA 720 Benchtop Chassis and TLA 7XM Expansion Chassis Service Manual 7 11...
Page 203: ...Replaceable Parts 7 16 TLA 720 Benchtop Chassis and TLA 7XM Expansion Chassis Service Manual...
Page 204: ...Index...
Page 205: ......
Page 211: ...Index Index 6 TLA 720 Benchtop Chassis and TLA 7XM Expansion Chassis Service Manual...
Page 212: ......
Page 213: ......