Adjustment Procedures
ADA400A Differential Preamplifier
49
Table 22: Equipment Settings for LF Adjustments
ADA400A
Setting
+ Input
DC
-- Input
Gnd
Gain
1
Upper Bandwidth
Full (1 MHz)
Oscilloscope
Setting
Vertical gain, Ch1
200 mV/div
Time base
20
s/div
Calibration Generator
Setting
Output Mode
Fast Rise, 10 kHz
2.
Connect dual input connector to ADA400A input.
3.
Connect 50
Ω
terminator to dual input connector.
4.
Connect a coaxial cable to calibration generator fast rise output to 50
Ω
terminator.
5.
Connect a coaxial cable from trigger out of calibration generator to Ch4 of
oscilloscope.
6.
Adjust amplitude for 5 divisions on screen.
7.
Adjust CV6 for flat response.
8.
Make the second equipment settings according to Table 23.
Table 23: Second Equipment Settings for LF Adjustments
ADA400A
Setting
+ Input
DC
-- Input
DC
Oscilloscope
Setting
Vertical gain
5 mV/div
Time base
20
s/div
9.
Adjust CV14 for minimum trace.
10.
Disconnect setup.
This completes the ADA400A adjustment procedures.