ADA400A Differential Preamplifier
29
Performance Verification
This section contains procedures for checking that the ADA400A Differential
Preamplifier meets warranted specifications.
Requirements for Performance
The performance limits of this procedure are valid with the following conditions:
H
The ADA400A Differential Preamplifier is in an environment that is within
the operating limits of temperature, altitude, humidity, and vibration
described in
Specifications.
H
The data line jumper located inside the ADA400A Differential Preamplifier.
is properly configured depending on the type of oscilloscope you connect to
the output. See the
Installation
section on page 2.
H
The ADA400A Differential Preamplifier has a period of at least a 20 minutes
to warm up prior to verification.
Equipment Required
Table 9 lists the equipment required to verify and adjust the ADA400A
Differential Preamplifier. The
Adjustment Procedures
section begins on page 41.
Table 9: Test Equipment
Item Number and
Description
Minimum Requirements
Example
Purpose
1.
Digitizing Oscilloscope
Bandwidth: > 5 MHz
Sensitivity: 1 mV/div
Auto Measurement: Frequency,
Amplitude
Input Impedance: 1 M
Ω
Must have TekProbe Level 2
Meet functional check with
scale factor coding (see page
2).
----
Supply power to and display
output of ADA400A
2.
Sine Wave Generator
Frequency Range: 1 Hz to
5 MHz
SG5030
CMRR checks and adjustments
3.
Digital Multimeter
Resolution:
≥
3.5 digits
Accuracy:
≤
0.5%
Resistance Range: 200
Ω
to
10 M
Ω
DM2510
Input resistance check
Summary of Contents for TEKFAX ADA400A
Page 4: ......
Page 8: ...Table of Contents iv ADA400A Differential Preamplifier...
Page 18: ...Getting Started 6 ADA400A Differential Preamplifier...
Page 50: ...Performance Verification 38 ADA400A Differential Preamplifier...
Page 52: ......
Page 62: ...Adjustment Procedures 50 ADA400A Differential Preamplifier...
Page 76: ...Replaceable Parts 64 ADA400A Differential Preamplifier...