Performance Verification
ADA400A Differential Preamplifier
35
2.
Connect a coaxial cable from sine wave generator trigger output to Ch4 of
the oscilloscope.
3.
Connect dual input connector to inputs of ADA400A.
4.
Connect 50
Ω
terminator to dual input connector.
5.
Connect two 10X attenuators to 50
Ω
terminator.
6.
Connect a coaxial cable from sine wave generator output to 10X attenuators.
7.
Adjust trigger level for a steady display.
8.
Adjust the amplitude of the sine wave generator for 5 divisions on screen.
9.
Remove the two 10X attenuators and connect the cable directly to the 50
Ω
terminator.
10.
Set ADA400A ‘--’ input to
DC
.
11.
Set oscilloscope to 1 mV/div.
12.
Measure signal is
≤
1 division (1 mV, 60 dB).
13.
Modify the equipment settings according to Table 13.
Table 13: Equipment Settings for X100, 10 kHz CMRR Checks
ADA400A
Setting
Upper Bandwidth
100 kHz
Oscilloscope
Setting
Vertical gain
2 mV/div
Time base
50
s/div
Sine Wave Generator
Setting
Output
10 kHz
14.
Set ‘--’ input to
GND
.
15.
Do steps 5 to 10.
16.
Set oscilloscope to 10
V/div. Measure trace is
<
3.2 divisions wide (32
V,
90 dB).
17.
Modify the equipment settings according to Table 14.
Summary of Contents for TEKFAX ADA400A
Page 4: ......
Page 8: ...Table of Contents iv ADA400A Differential Preamplifier...
Page 18: ...Getting Started 6 ADA400A Differential Preamplifier...
Page 50: ...Performance Verification 38 ADA400A Differential Preamplifier...
Page 52: ......
Page 62: ...Adjustment Procedures 50 ADA400A Differential Preamplifier...
Page 76: ...Replaceable Parts 64 ADA400A Differential Preamplifier...