Operating Basics
10
ADA400A Differential Preamplifier
Table 2: Controls, Connectors, and Indicators (Cont.)
Feature
Description
Upper Bandwidth Selector.
The Upper Bandwidth Selector reduces electrical
interference by limiting the frequency response of the amplifier to FULL (1 MHz),
100 kHz, 3 kHz, or 100 Hz.
Output Connector.
The output connector contains connections for the TEKPROBE
interface and signal output.
Determining the Effective Volts/Division
If your oscilloscope cannot read the scale factor information, see Table 3 to
determine the effective volts/division setting. The effective volts per division is
the volts per division setting of the oscilloscope divided by the gain setting of
the ADA400A Differential Preamplifier. For example, if the volts/division
setting on the oscilloscope is 500 mV/division and the gain setting is 10, the
effective volts/division is 500 divided by 10 or 50 mV.
Table 3: Effective Volts Per Division Settings
Effective Volts Per Division
Scope Volts/Div
Gain = .1
Gain = 1
Gain = 10
Gain = 100
1
10
1
100 m
10 m
500 m
5
500 m
50 m
5 m
200 m
2
200 m
20 m
2 m
100 m
1
100 m
10 m
1 m
50 m
500 m
50 m
5 m
500
20 m
200 m
20 m
2 m
200
10 m
100 m
10 m
1 m
100
5 m
50 m
5 m
500
50
2 m
20 m
2 m
200
20
1 m
10 m
1 m
100
10
Summary of Contents for TEKFAX ADA400A
Page 4: ......
Page 8: ...Table of Contents iv ADA400A Differential Preamplifier...
Page 18: ...Getting Started 6 ADA400A Differential Preamplifier...
Page 50: ...Performance Verification 38 ADA400A Differential Preamplifier...
Page 52: ......
Page 62: ...Adjustment Procedures 50 ADA400A Differential Preamplifier...
Page 76: ...Replaceable Parts 64 ADA400A Differential Preamplifier...