Command Descriptions
2–70
TDS Family Oscilloscope Programmer Manual
Vertical
CH<x>?
CH
<x>
?
CH1?
might return the string
:CH1:SCALE 10.0E-3;POSITION 0.0E+0;
OFFSET 0.0E+0;COUPLING DC;IMPEDANCE MEG;BANDWIDTH FULL
for chan-
nel 1.
CH<x>:BANdwidth
Sets or queries the bandwidth setting of the specified channel. This is equivalent
to setting Bandwidth in the Vertical menu.
Vertical
CH<x>:BANdwidth { TWEnty | HUNdred
(TDS 400A, 510A)
| TWOfifty
(TDS 500C, 600B & 700C)
| FULl }
CH<x>:BANdwidth?
TWOfifty
HUNdred
BANdwidth
<Space>
?
FULl
CH
:
<X>
TWEnty
TWEnty
sets the channel bandwidth to 20 MHz.
HUNdred
sets the channel bandwidth to 100 MHz (TDS 400A, 510A).
TWOfifty
sets the channel bandwidth to 250 MHz (TDS 500C, 600B, & 700C).
FULl
sets the channel bandwidth to the full bandwidth of the digitizing oscillo-
scope.
CH2:BANDWIDTH TWENTY
sets the bandwidth of channel 2 to 20 MHz.
Group
Syntax
Examples
Group
Syntax
Arguments
Examples