Command Descriptions
2–306
TDS Family Oscilloscope Programmer Manual
TRIGger:MAIn:PULse:GLItch:FILTer { ACCept | REJect }
TRIGger:MAIn:PULse:GLItch:FILTer?
ACCept
TRIGger:
:
PULse
MAIn
:
:
FILTer
GLItch
REJect
<Space>
?
ACCept
specifies that the digitizing oscilloscope will trigger only on pulses that
are narrower than the specified width when the main trigger type is set to pulse
glitch. The width is specified using TRIGger:MAIn:PULse:GLItch:WIDth
command.
REJect
specifies that the digitizing oscilloscope will trigger only on pulses that
are wider than the specified width when the main trigger type is set to pulse
glitch. The width is specified using TRIGger:MAIn:PULse:GLItch:WIDth
command.
TRIGGER:MAIN:PULSE:GLITCH:FILTER?
returns either
ACCept
or
REJect
, indicating whether glitches are filtered.
TRIGger:MAIn:PULse:GLItch:POLarity
Not on TDS 400A
Sets or queries the polarity for the main pulse glitch trigger. This command is
equivalent to selecting Polarity & Width in the Trigger menu.
Trigger
TRIGger:MAIn:PULse:GLItch:POLarity { EITher | NEGAtive |
POSITIVe }
TRIGger:MAIn:PULse:GLItch:POLarity?
Syntax
Arguments
Examples
Group
Syntax