Appendix A: Specifications
TDS 200-Series Digital Oscilloscope User Manual
95
Specifications (Cont.)
Vertical
Delta Volts Measure-
ment Accuracy, Aver-
age Acquisition Mode
Delta volts between any two
averages of
≥
16 waveforms
acquired under same setup
and ambient conditions
±
(3%
×
r 0.05 div)
P2100 passive probe
10X position
1X position
Bandwidth
DC to 100 MHz
DC to 7 MHz
Attenuation ratio
10:1
1:1
Compensation Range
18 pf-35 pf
All oscilloscopes with 1 M
W
input
Input resistance
10 M
W
±
2%
1 M
W
±
2%
Input capacitance
14.5 pf-17.5 pf
80 pf-110 pf
Maximum input
voltage
10X position
300 V
RMS
CAT I or 300 V DC CAT I
300 V
RMS
CAT II or 300 V DC CAT II
100 V
RMS
CAT III or 100 V DC CAT III
1X position
150 V
RMS
CAT I or 150 V DC CAT I
150 V
RMS
CAT II or 150 V DC CAT II
100 V
RMS
CAT III or 100 V DC CAT III