background image

 

 

 


 

Basics  

This section will help you become familiar with adjusting the vertical and horizontal controls, taking 
continuous or a single measurements and how to adjust the trigger. 

Autoset and Default setup 

Normally when you are measuring a signal you can press the 

Autoset

 button and the oscilloscope will 

automatically adjust the vertical and horizontal scale setting to an appropriate setting and display the 
peak to peak voltage, mean, period and frequency values. You can adjust this setting following the 
instructions below. If you ever get the oscilloscope into a setting where you can’t see the signal your 
measuring, you can press the 

Default Setup

 button to restore the oscilloscope to a predefined state. 

Note

: the default setup button does not affect the save / recall settings. 

Vertical Menu 

The vertical menu allows you to measure the voltage of an alternating current or signal. The screen has 
8 blocks going from the bottom of the screen to the top. The value of each square is set by the large 
control knob (

Scale

 knob) below the channel number button. You will normally adjust the vertical 

Scale

 

knob where you can see the largest signal on the screen without having the signal go off the screen.  

The position knob above the channel number in the vertical menu moves the entire signal up or down. 

By pressing the 

channel number

 

(1 or 2)

 button, you can bring up the 

Coupling

 softkey. This will 

normally be set to 

DC

. This means you will see the DC component of the signal with the AC component. 

By setting the 

Coupling

 softkey to 

AC

, you will remove any DC voltage from the signal you are 

measuring. 

The probes that come with the TBS1052B-EDU are x10 probes. This means a 10 volt peak to peak signal 
will come into the oscilloscope at 1 volt peak to peak. Because we know the signal is actually 10 times 
larger, the oscilloscope is defaulted to show the signal it receives as 10 times the actual input. If you use 
an oscilloscope probe that is not a times 10 probe, you will need to press the 

channel number

 

(1 or 2)

 

button then the 

Probe

 softkey. This will bring up the 

Attenuation

 softkey. Press the

 Attenuation

 softkey 

and use the 

Multipurpose

 wheel to select the correct probe attenuation. 

Horizontal Menu 

The horizontal menu adjusts the time parameters of the oscilloscope. There are 10 blocks going from 
the left side of the screen to the right side of the screen. The large 

Scale

 knob in the Horizontal menu 

will adjust each square to a different time scale, from 50 seconds to 5 nanoseconds. The position knob in 
the horizontal menu moves the entire signal left and right on the screen. Pushing the 

Position

 knob will 

center the signal to the middle of the screen. The 

Acquire

 button will allow you to take continuous 

samples or take an average of different sample sizes of the signal. 

Trigger Menu 

The signal on the screen of the oscilloscope is drawn from left to right many, many times a second. In 
order for the oscilloscope to draw a repeating signal that is steady (doesn’t drift or freeze), it needs to 

Summary of Contents for TBS1052B-EDU

Page 1: ...0 Tektronix TBS1052B EDU Oscilloscopes UTA Quick Guide Author Reviewer Date Revision Tige Kelley 01 10 2017 Draft...

Page 2: ...3 Probe Compensation 3 Basics 5 Vertical Menu 5 Horizontal Menu 5 Trigger Menu 5 Run Stop and Single Buttons 6 Firmware 6 Checking the firmware 6 Cursors 7 Measurements 8 Save Recall Screenshots CSV a...

Page 3: ...d above the channel button The volts per division or scale knob is located below the channel button 4 Trigger menu Press the Level knob to place the trigger at the 50 mark of the signal 5 Horizontal T...

Page 4: ...nput then twist the BNC connector right until the tabs are horizontal To remove the probe align the tabs vertically and pull the probe off the connector 3 Connect the Probe tip to the PROBE COMP outpu...

Page 5: ...4 Figure 3 Figure 4 Left is an under compensated probe and right is an over compensated probe Figure 5 A correctly compensated probe Grabber Tip Ground clip...

Page 6: ...lly be set to DC This means you will see the DC component of the signal with the AC component By setting the Coupling softkey to AC you will remove any DC voltage from the signal you are measuring The...

Page 7: ...of a computer Firmware is programmed into permanent memory of the device like an EEPROM Many times test equipment manufacturers will update their firmware to fix bug or add new features to a product I...

Page 8: ...use the Multipurpose wheel to move the cursor to the bottom of the signal your interested in The middle softkey will show you the difference in Amplitude between the two cursors 4 Selecting Time will...

Page 9: ...information from a USB drive on the oscilloscope Saving a BMP JPG to your USB drive 1 Insert a USB jump drive into the front of the oscilloscope 2 Press the Save Recall button 3 Press the Print Button...

Page 10: ...emory You can save up to 10 setups on the oscilloscope Select a setup number then press Save 3 If you wish to store the setup on a USB drive Insert a USB jump drive into the front of the oscilloscope...

Reviews:

Related manuals for TBS1052B-EDU