![Tektronix TAS 200 Instruction Manual Download Page 46](http://html1.mh-extra.com/html/tektronix/tas-200/tas-200_instruction-manual_1078183046.webp)
TAS 200 Series Adjustment Procedures
42
Bench Test Instruments and Handheld Oscilloscopes Basic Service
Probe Compensation
To locate the adjustment for the following procedure, refer to Figure 11. The
Power and High Voltage board occupies the bottom-left side of the instrument
below the CRT.
Instrument
front
Test points
ASTIG
VR604
SUB INTENSITY
VR603
SUB FOCUS
VR602
+12V ADJ
VR601
FREQ ADJ
VR701
Figure 11: Power and High Voltage board (viewed from the instrument bottom)
1.
Set up the oscilloscope as follows:
VERTICAL MODE
CH1
CH 1 VOLTS/DIV
0.5 V
CH 1 AC-DC
DC
HORIZONTAL SEC/DIV
0.2 ms
TRIGGER COUPLING
DC
TRIGGER SOURCE
CH 1
2.
Connect the probe to the CH 1 input connector
3.
Touch the probe tip to the PROBE COMP tab on the left side of the front
panel.
4.
Adjust the probe (compensation) for a flat-topped square wave on the
display.
5.
Adjust VR701 on the Power and High Voltage board until one cycle spans
five divisions. See Figure 11 for the adjustment location.
6.
Verify that the display amplitude is 2 V
p-p
±
2% and the duty-cycle ratio is
approximately 1 to 1.