TAS 200 Series Performance Verification
Bench Test Instruments and Handheld Oscilloscopes Basic Service
9
Set Up
Following a 20 minute warm-up period, preset the oscilloscope to the settings
listed below.
Table 12: Oscilloscope initial settings
Control
Setting
INTENSITY
Visible Display
CURSOR ON/OFF
Off
VERTICAL
POSITION
Midrange
MODE
CH1
VOLTS/DIV
10 mV
VARIABLE
CAL
GND
In
HORIZONTAL
SEC/DIV
10 ms
POSITION
Midrange
MAG
Off (Out)
X-Y
Off (Out)
SWEEP VARIABLE
CAL
TRIGGER
SLOPE
Positive (push switch out)
LEVEL
Midrange
MODE
AUTO
SOURCE
CH 1
COUPLING
DC