TAS 200 Series Performance Verification
8
Bench Test Instruments and Handheld Oscilloscopes Basic Service
Test Equipment
The performance verification procedures use external traceable test equipment to
directly check warranted characteristics.
Alternative test equipment must meet or exceed the intended minimum
requirements. If you substitute equipment, you may need to modify the
performance verification procedures.
NOTE
. Before beginning the performance verification procedures, warm up the
test equipment according to the manufacturer’s recommendations.
Table 11: Performance verification test equipment
Description
Minimum requirements
Example product
Leveled Sine Wave Generator
200 kHz to 250 MHz; variable amplitude
from 5 mV to 4 V
p-p
into 50
Ω
Wavetek 9100 Universal Calibration
System with Oscilloscope Calibration
M d l (O i
)
Time Mark Generator
Variable marker frequency from 10 ms to
10 ns; accuracy within 2 ppm
y
p
Module (Option 250)
Fluke 5500A Multi-product Calibrator with
DC Calibration Generator
DC voltage levels from 100 mV to 10 V
Fluke 5500A Multi product Calibrator with
Oscilloscope Calibration Option 5500A-SC
Digital Multimeter
35 V, 4 A,
±
0.1% Accuracy
Tektronix DMM914 Digital Multimeter
Termination (two required)
50
Ω,
female BNC input, male BNC output
Tektronix 011-0049-01
Cable, Coaxial
75
Ω,
male-to-male BNC connectors,
36 inch length
Tektronix 012-1338-00
Cable, Precision Coaxial
50
Ω,
male-to-male BNC connectors,
36 inch length
Tektronix 012-0482-00