TAS 200 Series Adjustment Procedures
38
Bench Test Instruments and Handheld Oscilloscopes Basic Service
HORIZONTAL SEC/DIV
1 ms
SWEEP UNCAL
Out
2.
Connect the time mark generator to the CH 1 input with a 50
Ω
coaxial cable
and 50
Ω
termination.
3.
Set up the generator to produce a 1 ms output.
4.
Adjust the generator output to produce a waveform two divisions in
amplitude on the display.
5.
Adjust VR405 for a one marker per division display over the center eight
divisions. See Figure 8 for the adjustment location.
Use the following procedure to adjust the 1
s and .1
s timing.
1.
Set up the oscilloscope as follows:
VERTICAL MODE
CH1
CH 1 VERTICAL POSITION
Midrange
HORIZONTAL SEC/DIV
1
s
HORIZONTAL SWP UNCAL
Out
2.
Connect the time mark generator to the CH 1 input with a 50
Ω
coaxial cable
and 50
Ω
termination.
3.
Set up the generator to produce a 1
s output.
4.
Adjust the generator output to produce a waveform one division in amplitude
on the display.
5.
Adjust VR404 for a one marker per division display. See Figure 8 for the
adjustment location.
6.
Set the generator to produce a .1
s output.
7.
Set the HORIZONTAL SEC/DIV control to .1
s.
8.
Adjust VR802 for a one marker per division display. See Figure 8 for the
adjustment location.
Use the following procedure to adjust the X-gain accuracy.
1.
Set up the oscilloscope as follows:
VERTICAL MODE
X-Y
CH 1 VOLTS/DIV
10 mV
CH 1 VARIABLE
CAL
HORIZONTAL X-Y
In
1
s and .1
s Timing
X-Gain Accuracy