Calibration
RSA2203A & RSA2208A User Manual
1- 25
Calibrating DC Offset (Option 05 Only)
The DC offset calibration cancels DC offset that appears at 0 Hz in the baseband
(DC to 20 MHz). When you change the amplitude setting and the DC offset is
too obvious, run the DC offset calibration.
When DC offset appears at 0 Hz in
the baseband, run the calibration.
Figure 1- 20: DC offset
To run the DC offset calibration, do the following:
1.
Press the
CAL
key on the front panel.
2.
Press the
Calibrate DC Offset
side key.
The calibration runs. It takes several seconds to complete the process.
Summary of Contents for RSA2200A
Page 6: ......
Page 8: ......
Page 25: ...Getting Started...
Page 26: ......
Page 53: ...Operating Basics...
Page 54: ......
Page 82: ...Functional Overview 2 28 RSA2203A RSA2208A User Manual...
Page 119: ...Reference...
Page 120: ......
Page 208: ...Setting Timing Parameters 3 88 RSA2203A RSA2208A User Manual...
Page 220: ...Trigger 3 100 RSA2203A RSA2208A User Manual...
Page 256: ...Display Line 3 136 RSA2203A RSA2208A User Manual...
Page 268: ...Marker Operation and Peak Search 3 148 RSA2203A RSA2208A User Manual...
Page 274: ...Selecting Input Source 3 154 RSA2203A RSA2208A User Manual...
Page 298: ...File Format 3 178 RSA2203A RSA2208A User Manual...
Page 302: ...Screen Copy 3 182 RSA2203A RSA2208A User Manual...
Page 303: ...Appendices...
Page 304: ......
Page 320: ...Appendix A Specifications A 16 RSA2203A RSA2208A User Manual...
Page 336: ...Appendix D Options and Accessories D 4 RSA2203A RSA2208A User Manual...
Page 337: ...Glossary...
Page 338: ......
Page 346: ...Glossary Glossary 8 RSA2203A RSA2208A User Manual...
Page 347: ...Index...
Page 348: ......