6.

For P77STFLXA or P77STCABL tips, connect the tips to the

fi

xture as

follows:

a.

Locate the Port1 plastic clamp on the

fi

xture and press down on the

spring-loaded clamp. Insert one of the probe tips in the direction shown in
the following

fi

gure, and release the clamp so that the tip locks in place.

Figure 2: Connecting the probe tip to the deskew

fi

xture

A green LED lights up under the probe tip clamp when the tip is properly
inserted.

b.

Repeat the above step for inserting the second probe tip into the Port2
plastic clamp.

c.

Verify that the green LEDs under both probe tips are on. If not,
double-check your connections.

7.

For P77BRWSR browser accessory, there are several pads on the deskew

fi

xture that can be contacted by the browser tips. Using these pads, two probes

with browser accessories or one probe with a solder tip and one probe with a
browser can be deskewed relative to one another.

The pads on the

fi

xture provide access to the same signals routed to the

clamps used to connect the solder tips. There are also ground pads located in
the same area for use when deskewing using a single-ended signal.

NOTE.

The signals routed to the pads are delayed by 30 ps relative to the signal

output at the solder tip clamps. There is a 5.7 ps time difference between the

fi

rst row of pads and the second row.

P77DESKEW Fixture Instructions

3

Summary of Contents for P77DESKEW

Page 1: ...your probe user manual for the anti static information For best performance you should check the calibration status of the oscilloscope before performing the deskew procedure If the oscilloscope requi...

Page 2: ...ey will be used 3 Connect the TekFlex tips to the ends of the main probe cables 4 For single ended inputs connect an SMA cable from a single source such as the FAST EDGE output connector on the oscill...

Page 3: ...under both probe tips are on If not double check your connections 7 For P77BRWSR browser accessory there are several pads on the deskew fixture that can be contacted by the browser tips Using these pa...

Page 4: ...on the B Port4 side of fixture board edge located between the Port1 and Port2 clamps Press the second tip against one of the ground pads on the right side of the row Figure 3 Connection points for the...

Page 5: ...l to use as the reference channel Channel 1 works well for this purpose 9 Select Deskew from the Vertical menu 10 Select the reference channel button and set the Deskew to 0 0 s 11 Set the acquisition...

Page 6: ...Select the button for the second channel that will be matched in time to the reference channel NOTE Do the next step at a signal amplitude within the same attenuator range vertical scale as your plann...

Page 7: ...its signal aligns with that of the reference channel When the signals are aligned they should look similar to the following figure Figure 6 Aligned signals no delay 19 Repeat the preceding steps for...

Page 8: ...ignal output at the solder tip clamps There is a 5 7 ps time difference between the first row of pads and the second row Environmental information Equipment Recycling This product complies with the Eu...

Reviews: